Optimization of axial-pump pressure sensitivity for a continuous-flow total artificial heart.
MedLine Citation:
PMID:  20133164     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: In this study, we describe the potential advantages of a continuous-flow total artificial heart (CFTAH) comprising two small, non-pulsatile pumps with optimized responsiveness to the pressure gradient. METHODS: We modified a MicroMed DeBakey axial-flow pump by increasing its inducer-impeller inlet angle, thereby increasing its pressure responsivity. We obtained the in vitro pressure gradient response and compared it with those of the clinically used, unmodified MicroMed DeBakey pump, Jarvik 2000 FlowMaker and HeartMate II. RESULTS: The modified pump showed an increased response to changes in the pressure gradient at pump flow rates of between 2 and 4 liters/min. The maximum pressure responsivity of the modified pump was 2.5 liters/min/mm Hg; the corresponding maximum responsivities of the Jarvik 2000, HeartMate II and MicroMed DeBakey ventricular assist devices (VADs) were 0.12, 0.09 and 0.38 liters/min/mm Hg, respectively. CONCLUSIONS: Because of the inherent properties of non-pulsatile pumps, the CFTAH may potentially respond to changes in inflow and outflow pressures while maintaining physiologic flow rates sufficient for normal daily activity. In addition, the hemodynamic interplay between the two optimized pumps should allow a physiologic response to normal flow imbalances between the pulmonary and systemic circulations. Improved responsiveness to inflow pressure may further simplify and improve the CFTAH and affect its potential clinical use as a meaningful therapy for terminal heart failure.
Authors:
O H Frazier; Hassan A Khalil; Robert J Benkowski; William E Cohn
Related Documents :
3172364 - Pressure-controlled hydraulic dilation of the ureter: "one-step" ureteroscopy.
17534694 - Non-invasive low frequency vibration as a potential emergency adjunctive treatment for ...
1760514 - Heart rate and cardiac energetics.
10463494 - Defining pulsatile perfusion: quantification in terms of energy equivalent pressure.
24697434 - Ignition experiments and models of a plastic bonded explosive (pbx 9502).
3423394 - Difference between intracarotid and intravenous infusions of angiotensin ii on barorefl...
Publication Detail:
Type:  Journal Article; Research Support, N.I.H., Extramural     Date:  2010-02-04
Journal Detail:
Title:  The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation     Volume:  29     ISSN:  1557-3117     ISO Abbreviation:  J. Heart Lung Transplant.     Publication Date:  2010 Jun 
Date Detail:
Created Date:  2010-05-25     Completed Date:  2010-09-23     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9102703     Medline TA:  J Heart Lung Transplant     Country:  United States    
Other Details:
Languages:  eng     Pagination:  687-91     Citation Subset:  IM    
Copyright Information:
Copyright 2010 International Society for Heart and Lung Transplantation. Published by Elsevier Inc. All rights reserved.
Affiliation:
Department of Cardiovascular Surgery, Texas Heart Institute at St. Luke's Episcopal Hospital, Houston, Texas 77225-0345, USA. lschwenke@heart.thi.tmc.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Pressure*
Coronary Circulation*
Equipment Design
Heart, Artificial*
Heart-Assist Devices
Hemodynamics
Humans
Pulsatile Flow
Grant Support
ID/Acronym/Agency:
R01HL085054/HL/NHLBI NIH HHS

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Azithromycin is associated with increased survival in lung transplant recipients with bronchiolitis ...
Next Document:  Paracorporeal biventricular mechanical circulatory support for more than 4 years.