Novel intracorporeal vascular knot applier for large vessel ligation in laparoscopic surgery: experience in pigs and preliminary clinical use.
MedLine Citation:
PMID:  12689398     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND AND PURPOSE: We designed an intracorporeal knot applier for the ligation of large vessels laparoscopically. The aim of this study was to assess the efficacy and efficiency of this method to control large vessels in pigs. MATERIALS AND METHODS: Six pigs with mean body weight of 28 kg (range 26-30 kg) were enrolled. The standard transperitoneal laparoscopic set-up was established with 15 mm Hg of pneumoperionteum. The right renal hilum was mobilized, and the renal artery and renal vein were separated. A 5-cm segment of inferior vena cava with the right renal vein was skeletonized laparoscopically. Assisted by the novel knot applier, 3-0 silk was used to ligate the renal artery, renal vein, and inferior vena cava. The time required to accomplish these ligatures was recorded. Two intracorporeal knots were applied 5 mm apart, and the vessels were divided between the knots. The knot strength was assessed by measuring the maximal intraluminal pressure required to blast out the knot using gas insufflation. RESULTS: The diameters (mean +/- SD) of the porcine inferior vena cava, renal artery, and renal vein were 11.1 +/- 0.7, 2.9 +/- 0.3, and 4.9 +/- 0.3 mm, respectively. The knotting time was 58.9 +/- 8.6, 57.5 +/- 6.6, and 59.1 +/- 6.3 seconds for these vessels, respectively. No bleeding occurred when the vessels were divided. The average pressure needed to withstand the blast out of the ligature on the inferior vena cava, renal artery, and renal vein was 213.3 +/- 32.7 (range 165-250), 227.5 +/- 14.7 (range 210-250), and 210.8 +/- 21.5 (range 180-240) mm Hg, respectively. CONCLUSIONS: The novel intracorporeal knot applier was easy to use when making a ligature on vessels of different sizes and different wall thicknesses. The time required for making a ligature was short, and the knot was secure enough to withstand a vascular pressure higher than ordinary blood pressure. The laparoscopic device is inexpensive with great potential of being used as a secure vascular ligature provider.
Authors:
Allen W Chiu; Sheng-Tang Wu; Guang-Huan Sun; Catherine J Liu; Jyh-Jann Lin; Chiung-Ying Tseng; Jen-Tai Lin
Related Documents :
6484298 - Effects of volume expansion with fluosol-da on renal function of the nonhuman primate.
2949628 - Effect of reduction in renal artery pressure on atrial natriuretic peptide-induced natr...
6994698 - Hemofiltration for the treatment of hypertensions associated with end-stage renal failure.
1093748 - Primary role of renal homografts in setting chronic blood pressure levels in rats.
7836228 - Atrial septal defect blunts the impairment of left ventricular function during the muel...
11131248 - Cardiopulmonary changes during laparoscopy and vessel injury: comparison of co2 and hel...
Publication Detail:
Type:  Evaluation Studies; Journal Article    
Journal Detail:
Title:  Journal of endourology / Endourological Society     Volume:  17     ISSN:  0892-7790     ISO Abbreviation:  J. Endourol.     Publication Date:  2003 Mar 
Date Detail:
Created Date:  2003-04-11     Completed Date:  2003-06-25     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  8807503     Medline TA:  J Endourol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  73-8     Citation Subset:  IM    
Affiliation:
Division of Urology, Department of Surgery, Chi Mei Medical Center, Tainan, Taiwan, ROC. chimeigu@venus.seed.net.tw
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Hemostasis, Surgical / instrumentation*,  methods
Humans
Laparoscopy / methods*
Ligation / instrumentation,  methods
Pneumoperitoneum, Artificial
Renal Artery / surgery
Renal Veins / surgery
Surgical Instruments
Swine
Vena Cava, Inferior / surgery

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Novel approach to minimizing trocar sites during challenging hand-assisted laparoscopic surgery util...
Next Document:  Rapid communication: laparoscopic Anderson-Hynes dismembered pyeloplasty using the da Vinci robot: t...