| Noninvasive continuous beat-to-beat radial artery pressure via TL-200 applanation tonometry. | |
| | |
MedLine Citation:
|
PMID: 22258303 Owner: NLM Status: Publisher |
Abstract/OtherAbstract:
|
The Tensys TL-200(®) noninvasive beat-to-beat blood pressure (BP) monitor displays continuous radial artery waveform as well as systolic, mean and diastolic BP from a pressure sensor directly over the radial artery at the wrist. It locates the site of maximal radial pulse signal, determines mean BP from maximal pulse waveform amplitude at optimal artery compression and then derives systolic and diastolic BP. We performed a cross-sectional study of TL-200 BP comparisons with contralateral invasive radial artery (A-Line) BP values in 19 subjects during an average 2.5 h of general anesthesia for a wide range of surgical procedures. Two hundred and fifty random sample pairs/patient resulted in 4,747 systolic, mean and diastolic BP pairs for analysis. A-Line BP ranged from 29 mm Hg diastolic to 211 mm Hg systolic, and heart rate varied between 38 and 210 beats/min. Bland-Altman analysis showed an average 2.3 mm Hg TL-200 versus A-Line systolic BP bias and limits of agreement (1.96 SD) were ± 15.3 mm Hg. Mean BP showed a 2.3 mm Hg TL-200 bias and ± 11.7 mm Hg limits of agreement, while diastolic BP showed a 1.7 mm Hg bias and ± 12.3 mm Hg limits of agreement. Coefficients of determination for TL-200 and A-Line BP regression were r² = 0.86 for systolic, r² = 0.86 for mean, and r² = 80 for diastolic BP, respectively, with no apparent change in correlation at low or high BP. Bland-Altman analysis suggested satisfactory agreement between TL-200 noninvasive beat-to-beat BP and invasive A-Line BP. Paired TL-200/A-Line BP comparisons showed a high coefficient of determination. |
| | |
Authors:
|
Ron Dueck; Oliver Goedje; Paul Clopton |
Related Documents
:
|
11834673 - Effects of partial liquid ventilation on unilateral lung injury in dogs. 11580113 - Reversible temperature-sensitive alterations in lung fluid balance. 9696073 - Preoperative volume expansion improves tolerance to carotid artery cross-clamping durin... |
Publication Detail:
|
Type: JOURNAL ARTICLE Date: 2012-1-18 |
Journal Detail:
|
Title: Journal of clinical monitoring and computing Volume: - ISSN: 1573-2614 ISO Abbreviation: - Publication Date: 2012 Jan |
Date Detail:
|
Created Date: 2012-1-19 Completed Date: - Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 9806357 Medline TA: J Clin Monit Comput Country: - |
Other Details:
|
Languages: ENG Pagination: - Citation Subset: - |
Affiliation:
|
Anesthesiology Service, Veterans Affairs San Diego Healthcare System, San Diego, CA, USA, rdueck@vapop.ucsd.edu. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
|
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Collaborative eye tracking: a potential training tool in laparoscopic surgery.
Next Document: Effect of intravenous hydralazine infusion on maternal plasma nitric oxide levels in gestations comp...