Document Detail

Molecular beacon probes-base multiplex NASBA Real-time for detection of HIV-1 and HCV.
Jump to Full Text
MedLine Citation:
PMID:  22973469     Owner:  NLM     Status:  PubMed-not-MEDLINE    
Abstract/OtherAbstract:
BACKGROUND AND OBJECTIVES: Developed in 1991, nucleic acid sequence-based amplification (NASBA) has been introduced as a rapid molecular diagnostic technique, where it has been shown to give quicker results than PCR, and it can also be more sensitive. This paper describes the development of a molecular beacon-based multiplex NASBA assay for simultaneous detection of HIV-1 and HCV in plasma samples.
MATERIALS AND METHODS: A well-conserved region in the HIV-1 pol gene and 5'-NCR of HCV genome were used for primers and molecular beacon design. The performance features of HCV/HIV-1 multiplex NASBA assay including analytical sensitivity and specificity, clinical sensitivity and clinical specificity were evaluated.
RESULTS: The analysis of scalar concentrations of the samples indicated that the limit of quantification of the assay was <1000 copies/ml for HIV-1 and <500 copies/ml for HCV with 95% confidence interval. Multiplex NASBA assay showed a 98% sensitivity and 100% specificity. The analytical specificity study with BLAST software demonstrated that the primers do not attach to any other sequences except for that of HIV-1 or HCV. The primers and molecular beacon probes detected all HCV genotypes and all major variants of HIV-1.
CONCLUSION: This method may represent a relatively inexpensive isothermal method for detection of HIV-1/HCV co-infection in monitoring of patients.
Authors:
S Mohammadi-Yeganeh; M Paryan; S Mirab Samiee; V Kia; H Rezvan
Related Documents :
24188659 - Policy implementation of methadone maintenance treatment and hiv infection: evidence fr...
11364049 - Hospital volunteer can't sue employer for botched hiv test.
24986599 - Realizing the potential for hiv self-testing.
25017549 - Bringing new hiv infections to zero - opportunities and challenges offered by antiretro...
24028699 - Shifting the paradigm: using hiv surveillance data as a foundation for improving hiv ca...
7866169 - Outbreak of hiv infection in a scottish prison.
17935229 - Distinct hepatitis c virus core and f protein quasispecies in tumoral and nontumoral he...
11588269 - An outbreak of early syphilis: cases from north manchester general hospital.
19768189 - Molecular pathogenesis of liver fibrosis.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Iranian journal of microbiology     Volume:  4     ISSN:  2008-4447     ISO Abbreviation:  Iran J Microbiol     Publication Date:  2012 Jun 
Date Detail:
Created Date:  2012-09-13     Completed Date:  2012-09-14     Revised Date:  2013-05-30    
Medline Journal Info:
Nlm Unique ID:  101518404     Medline TA:  Iran J Microbiol     Country:  Iran    
Other Details:
Languages:  eng     Pagination:  47-54     Citation Subset:  -    
Affiliation:
Biotechnology Research Center, Pasteur Institute of Iran, Tehran, Iran.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Iran J Microbiol
Journal ID (iso-abbrev): Iran J Microbiol
Journal ID (publisher-id): IJM
ISSN: 2008-3289
ISSN: 2008-4447
Publisher: Tehran University of Medical Sciences
Article Information
Download PDF
© 2012 Iranian Society of Microbiology & Tehran University of Medical Sciences
open-access:
Received Month: 2 Year: 2012
Accepted Month: 4 Year: 2012
Print publication date: Month: 6 Year: 2012
Volume: 4 Issue: 2
First Page: 47 Last Page: 54
ID: 3434641
PubMed Id: 22973469
Publisher Id: IJM-4-47

Molecular beacon probes–base multiplex NASBA Real-time for detection of HIV-1 and HCV
S Mohammadi-Yeganeh12
M Paryan12*
S Mirab Samiee34
V Kia5
H Rezvan3
1Biotechnology Research Center, Pasteur Institute of Iran, Tehran, Iran
2Molecular Biology and Genetic Engineering Department, Stem Cell Technology Research Center, Tehran, Iran
3Day General Hospital Laboratory, Tehran, Iran
4Food and Drug Laboratory Research Center, Ministry of Health and Medical Education, Tehran, Iran
5Department of Medical Biotechnology, Tarbiat Modares University, Tehran, Iran
* Corresponding author: Mahdi Paryan, Address: Biotechnology Research Center, Pasteur Institute of Iran, Tehran, Iran. Tel/ Fax: + 98 21 22351460. E-mail: mparyan@gmail.com

INTRODUCTION

Human immunodeficiency virus-1 (HIV-1) and hepatitis C virus (HCV) can transmit through common routes such as mother to child in pregnant women, injection drug abuse, and blood or blood products (14). Approaches to reduce transmission of these blood-borne agents from donor to recipient are stringent donor selection and performing screening tests to detect anti-HIV and anti-HCV seroreactivity in donors (5, 6). However, during a window period of these viruses when most serological tests are negative, infection cannot be detected in donors with newly acquired infection. Therefore, although sensitive serological tests have been developed effectively and successfully, there is still a residual risk of viral infection, due to the possibility of the HIV or HCV transmission either during the so-called window period, when viremia precedes seroreactivity, or in the presence of different viral serotypes, where the serological tests may be ineffective (57).

In order to reduce this residual risk, application of molecular techniques is recommended and is in use in several countries. However, molecular diagnostic technology has a major disadvantage namely high expense (1, 7). In this respect multiplex assays are developed to reduce the cost as well as running time (1, 710). A number of platforms of isothermal nucleic acid amplification have been developed and introduced in the past 18 years (11, 12).

NASBA was selected since it is a simple and rapid alternative for nucleic acid amplification in case of RNA viruses. In this technique RNA is amplified to a billion fold in around two hours (Fig. 1) (13). It is an isothermal process and therefore no advanced equipment such as thermal cycler is required(14). The NASBA process requires fewer cycle than polymerase chain reaction (PCR) to yield a desired amplification. When using PCR, the number of molecules doubles each step, therefore it requires about 20 cycles to amplify the template to a million-fold(14). With NASBA, however, 10-100 copies of RNA are generated in each transcription step, thus only four to five cycles are required to achieve a similar amplification. As a result, both the total incubation time and the overall error frequency are lowered with NASBA. Errors that are intrinsic in some enzymatic activities (e.g. reverse transcriptase) are cumulative, and therefore fewer cycles should reduce the probability of these errors occurring (14).The objective of the present study was to develop a sensitive and rapid multiplex real-time nucleic acid sequence-based amplification (NASBA) assay to simultaneously detect HIV-1 and HCV genomes, using molecular beacon probes. This multiplex real-time NASBA was assessed using HIV-1 and HCV specific oligonucleotides that encompass RNA sequences in viral conserved regions of genome in order to yield primers able to detect all the HCV genotypes and HIV-1 major subtypes.


MATERIALS AND METHODS
Clinical samples

Fifty positive plasma samples, including 30 positive samples from co-infected HIV-1 and HCV patients and 20 HIV-1 and HCV negative plasma samples spiked with dilutions of transcribed RNA, were tested to verify the efficacy of the NASBA real-time assay. Twenty healthy plasma samples approved negative for HCV and HIV-1 RNA by immunoassay method, were also used to assess the specificity.

Primers and probes design

We designed two sets of primers and molecular beacon probes on pol gene and 5'NCR region of HIV-1 and HCV viruses, respectively. All the conserved sequences were selected by alignment of maximum subtype and genotype sequences retrieved from the National Center for Biotechnology Information (NCBI, Bethesda, MD) GenBank database. Sequence alignments were performed using Mega4 software (Fig. 2 and 3). To ensure the multiplexed primers would have no mis-combinations we used Beacon Designer 7.0 (Premier Biosoft, Palo Alto, CA) software to confirm the sequencing reaction. The secondary structure of primers and target sequences were assessed by Mfold (http://mfold.rna.albany.edu/?q=mfold/RNA-Folding-Form). The chosen primers amplify a 179-base pair fragment of pol gene and 241-base pair fragment of 5'Non coding region (5'NCR). For distinction between the fluorescent signal of HIV-1 and HCV, sequence-specific probes were used with two different reporter dyes (Table 1).

Viral RNA extraction from plasma

Viral RNA extraction were carried out using QiaAmp RNA mini kit (Qiagen) according to the manufacturer's manual and eluted with 50µl of nuclease-free water and stored at -80 °C until use.

Cloning of HIV-1 and HCV fragments

The amplified fragments were gel purified (Qiagen, Germany) and cloned in PTZ57 R/A cloning vector as the manufacturer's instructions (Fermentas, Germany).

In-vitro transcription for standard preparation

The recombinant plasmid was linearised with EcoRI enzyme (Fermentas, Germany). In-vitro transcription was performed in a final volume of 50 µl using the recombinant plasmid and T7 RNA Polymerase (Fermentas, Germany), in presence of 2 mM each of the ribonucleoside 5'triphosphates, 10 µl transcription buffer, 50 U RNase inhibitor and 30 U T7 RNA Polymerase in a 50 µl reaction and tube, which was incubated for 120 min at 37°C. After transcription, 5U (2U per 1 µg of DNA used) of RNase free DNase (Fermentas, Germany) was added and the tubes incubated for 30 min at 37°C to degrade the template DNA. The transcribed RNA was purified by Trizol extraction method. The integrity of RNA was checked on a 2% formaldehyde agarose gel by electrophoresis (Fig. 4) and quantitated by measuring the optical density (OD) at 260 nm.

Multiplex NASBA Real-time

The multiplex procedure was carried out with final reaction mixture volume of 25 µl containing 40 mM Tris-HCl (pH 8.5), 50 mM KCl, 12 mM MgCl2, 1 mM each of the dNTPs, 2 mM each of the ribonucleoside 5'triphosphates, 10 mM dithiothreitol, and 15% (vol/vol) dimethyl sulfoxide. 0.4 µM of each primer, 0.2 µM each probe was used for the multiplex NASBA. 5 µl of purified RNA extracted from each sample was added to 18 µl of multiplex amplification mixture in a 0.2-ml microcentrifuge tube, which was incubated for 5 min at 65°C in order to disrupt secondary structures in the target RNA. Each tube was immediately cooled to 41°C for 5 min, after which 2 µl of an enzyme mixture containing 2.6 µg of bovine serum albumin (in 50% glycerol; Roche Diagnostics Corp., Indianapolis, Ind.), 40U of T7 RNA polymerase, 8 U of avian myeloblastosis virus reverse transcriptase, 0.2 U of RNase H and 12.5 U of RNasin (All enzymes were purchased from Fermentas, except AMV-reverse transcriptase from Roche) were added.

Microtubes were transferred to a Rotor-Gene 3000 (Corbett Research, Australia). Development of fluorescence was followed in closed tubes for 90 min at 41°C. Fluorescence intensity data were recorded every minute of the NASBA reaction. Moreover, NASBA amplification products were analyzed at the end of the amplification step, by electrophoresis of 5 µl of reaction on a 2% agarose gel containing 0.5 mg/ml EtBr. Gels were run at 100 V for 30 min in TAE buffer (40 mM Tris-acetate, 1 mM EDTA, pH 8.0). Bands were visualized by UV excitation and photographed.

Sensitivity and specificity

To determine the analytical specificity, we performed BLAST using NCBI BLAST program. To determine the clinical specificity, some viruses(e.g. HBV, HTLV-1, B19, HSV-1, HSV-2, HHV-6, HHV8, HCMV and EBV) and 20 negative samples were tested. These samples were previously tested for HIV-1 and HCV using immunoassay method. To determine the analytical sensitivity, the transcribed RNA mentioned above was used. A serial dilution of each standard from 104-50 copies/ml was prepared. We used the instruction in URI Genomics & Sequencing Center (http://www.uri.edu/research/gsc/resources/cndna.html) to convert optical density of transcribed RNA to copy number. To estimate the minimal detection limit, 10-fold serial dilutions of viral RNA samples were made in DNase and RNase free water. To assess the clinical sensitivity, fifty positive samples were tested to verify the efficacy of the NASBA real-time assay.


RESULTS
Multiplex real-time NASBA for detection of HIV-1 and HCV

Two molecular beacons with different fluorophore labels were designed to distinguish between HIV-1 and HCV viruses (Fig. 5a and b). These probes recognized the 5'NCR region of HCV and pol gene of HIV-1, they showed successful specific hybridization with their respective target sequences during NASBA amplification. For each virus, amplification led to a rise in fluorescence that was characteristic of the labeled fluorophores attached to their respective 5'NCR region or pol gene sequence. The sequence of the beacons and the amplification primers were selected due to showing better sequence conservation in comparison with other regions or genes. Nevertheless, the few mutations observed in some genotypes of HCV and subtypes of HIV-1 were not located at critical positions of the primers and probe, such as the 3′ and 5′ end of the primers and probe, respectively.

As shown in (Fig. 6), agarose gel electrophoresis confirmed the exact length of the amplicons: two bands of 179bp and 241bp, specific for HIV-1 and HCV, respectively, were detected in all of the co-infected samples analyzed.

Analytical sensitivity and specificity assay

For HIV-1/HCV viruses dilutions from transcribed RNA were tested in multiplex real-time NASBA assay from 104 to 50 copies/ ml. To estimate the analytical sensitivity, serial dilutions of each virus were prepared. The analytical sensitivity of assay was 1000 copies/ml for HIV-1 and 500copies/ ml for HCV. Moreover, the limit of quantification assay was <1000 copies/ml for HIV-1 and <500 copies/ ml for HCV with 95% confidence interval (Table 2).

Cross-reactivities were not observed when HBV, HTLV-1, B19, HSV-1, HSV-2, HHV-6, HHV8, HCMV and EBV samples were tested by the multiplex Real-time NASBA assay (data not shown) that confirmed the specificity.

Clinical sensitivity and specificity

49 samples out of 50 positive samples were positive for both viruses and one sample was positive only for HCV. Multiplex real-time NASBA assay showed a 98% sensitivity and 100% specificity.


DISCUSSION

The exigency for molecular diagnostic tests for blood donations in addition to some of the currently licensed serological methods is obvious because of the ability of these assays to remarkably reduce the window period during which serological methods fail to detect the infection (1517). A negative result by serological methods may be a false-negative result if the donor has been recently infected.

Conventional diagnostic methods such as immunological and serological tests have lower sensitivity than molecular techniques such as the PCR (18, 19). Nevertheless, the cost and necessity for technical expertise makes application of PCR more difficult for laboratories in developing countries and rural areas(18).

NASBA is a homogeneous and direct RNA amplification process which has been shown previously to have a sensitivity comparable to PCR(18). No thermocycler is required and a standardized constant reaction temperature is applied for highly reliable amplification of a RNA target. Due to reduction in expenses, the multiplex HIV-1/HCV NASBA assay could also be applied as a confirmatory technique for detection of these viruses.

Previous studies suggested molecular tests as a replacement for commercially available seroconversion panels (20, 21). Although there is not any report available for HIV-1 /HCV multiplex NASBA assay, this technique wildly used for detection of each virus.

The commercial kits for HIV-1 detection are accessible. The NucliSens® HIV-1 QT is an in vitro nucleic acid sequence based amplification test for the quantitation of HIV-1 RNA in human plasma. The test can quantitate HIV-1 RNA over the range of 176 to 3.47X106 copies/mL. The test is intended for use in conjunction with clinical presentation and other laboratory markers of disease progression for prognostic assessment of HIV-1 infected patients, and for monitoring the effects of anti-retroviral therapy by serial measurements of plasma HIV-1 RNA for pediatric and adult patients with baseline viral loads greater than 93,000 and 28,000 copies of HIV-1 viral RNA/ml, respectively(www.biomerieux-diagnostics.com).

Workenesh Ayele, et al. [2004] developed a Nucleic Acid Sequence-Based Amplification assay that used gag-based molecular beacons to distinguish between HIV-1 subtype C and C’ infections in Ethiopia. This assay is country specific for HIV-1 subtype C epidemics and it will contribute to characterizing the circulating viruses in this population, thereby generating the information necessary for the development of a potential efficacious HIV-1 appropriate vaccine(22).

In another experiment Rute Antunes, et al. evaluated the clinical sensitivitiy of three viral load assays with plasma samples from a pediatric population predominantly infected with HIV-1 subtype G and BG recombinant forms. To determine the viral load, AMPLICOR HIV-1 Monitor Test 1.5, Nuclisens HIV-1 QT, and Quantiplex HIV RNA 3.0 (bDNA) were investigate. AMPLICOR v1.5 and Quantiplex v3.0 detected all positive samples with a good correlation. Thirty-eight out of 61 specimens containing HIV-1 subtype B, G, or recombinant BG, could not detected by Nuclisens HIV-1 QT (23).

We have described a multiplex assay in which molecular beacons are used to detect amplicons in real-time NASBA under homogeneous conditions. In this assay, the hairpin stem of the molecular beacons keeps the quencher and the fluorophore at such close proximity to each other that quenching occurs as a result of physical contact between the two labeled moieties.

The multiplex NASBA assay for detection of HIV-1/HCV has advantages with respect to time and cost in comparison to assays designed with a single-target assay. Using multiplex aligning sequences in nucleotide database of NCBI (www.ncbi.nlm.nih.gov/nucleotide), two sets of primers and molecular beacon probes on the most conserved region of pol gene of HIV-1 and 5'NCR region of HCV were designed.

The data presented in this paper demonstrated that these primers and probes match the well-conserved sequences among different HCV genotypes as well as major HIV-1 subtypes. In fact, genotype independent detection of any virus such as HIV-1 and HCV is reliant upon locating the most conserved region of viruses genome by using a complete databank of aligned sequences for genome of both viruses.

This study demonstrates that the multiplex real-time NASBA assay using the appropriately designed molecular beacon probes has 98% clinical sensitivity and 100% specificity for the detection of HIV-1/HCV from plasma samples. Based on the analytical sensitivity data, the multiplex real-time NASBA assay showed significantly lower limit of quantification for HCV (<500 copies/ml) than that of HIV-1 (<1000 copies/ml) with 95% confidence interval and no cross-reactions were observed when other blood-borne viruses were tested by the multiplex real-time NASBA assay.


ACKNOWLEDGEMENT

The authors thank the workers of the Day General Hospital, Tehran, Iran for their cooperation.


REFERENCES
1. Gibellini D,Gardini F,Vitone F,Schiavone P,Furlini G,Re MC. Simultaneous detection of HCV and HIV-1 by SYBR Green real time multiplex RT-PCR technique in plasma samplesMol Cell ProbesYear: 2006203-422322916537101
2. Alter MJ,Kruszon-Moran D,Nainan OV,McQuillan GM,Gao F,Moyer LA,et al. The prevalence of hepatitis C virus infection in the United States, 1988 through 1994N Engl J MedYear: 1999341556562 19. 10451460
3. Alter MJ,Margolis HS,Krawczynski K,Judson FN,Mares A,Alexander WJ,et al. The natural history of community-acquired hepatitis C in the United States. The Sentinel Counties Chronic non-A, non-B Hepatitis Study TeamN Engl J MedYear: 19923272718991905 31. 1280771
4. Sherman KE,Rouster SD,Chung RT,Rajicic N. Hepatitis C Virus prevalence among patients infected with Human Immunodeficiency Virus: a cross-sectional analysis of the US adult AIDS Clinical Trials GroupClin Infect DisYear: 20023483183711833007
5. Ellingson K,Seem D,Nowicki M,Strong DM,Kuehnert MJ. Estimated risk of human immunodeficiency virus and hepatitis C virus infection among potential organ donors from 17 organ procurement organizations in the United StatesAm J TransplantYear: 2011111201120821645253
6. Ison MG,Llata E,Conover CS,Friedewald JJ,Gerber SI,Grigoryan A,et al. Transmission of human immunodeficiency virus and hepatitis C virus from an organ donor to four transplant recipientsAm J TransplantYear: 2011111218122521645254
7. Busch MP,Kleinman SH,Jackson B,Stramer SL,Hewlett I,Preston S. Nucleic acid amplification testing of blood donors for transfusion-transmitted infectious diseases: Report of the Interorganizational Task Force on Nucleic Acid Amplification Testing of Blood DonorsTransfusionYear: 20004014315910685998
8. Meng Q,Wong C,Rangachari A,Tamatsukuri S,Sasaki M,Fiss E,et al. Automated multiplex assay system for simultaneous detection of hepatitis B virus DNA, hepatitis C virus RNA, and human immunodeficiency virus type 1 RNAJ Clin MicrobiolYear: 2001392937294511474017
9. Mine H,Emura H,Miyamoto M,Tomono T,Minegishi K,Murokawa H,et al. High throughput screening of 16 million serologically negative blood donors for hepatitis B virus, hepatitis C virus and human immunodeficiency virus type-1 by nucleic acid amplification testing with specific and sensitive multiplex reagent in JapanJ Virol MethodsYear: 20031121-214515112951223
10. Pawlotsky JM. Use and interpretation of virological tests for hepatitis CHepatologyYear: 2002365 Suppl 1S657312407578
11. Tong Y,Lemieux B,Kong H. Multiple strategies to improve sensitivity, speed and robustness of isothermal nucleic acid amplification for rapid pathogen detectionBMC BiotechnolYear: 2011115021569347
12. Hill CS. Molecular diagnostic testing for infectious diseases using TMA technologyExpert Rev Mol DiagnYear: 20011444545511901859
13. Polstra AM,Goudsmit J,Cornelissen M. Development of real-time NASBA assays with molecular beacon detection to quantify mRNA coding for HHV-8 lytic and latent genesBMC Infect DisYear: 200221812207829
14. Compton J. Nucleic acid sequence-based amplificationNature.Year: 1991Month: 3 Day: 7 35063139121706072
15. Allain JP. Will genome detection replace serology in blood screening for microbial agents?Baillieres Best Pract Res Clin HaematolYear: 20001361562911102280
16. Giachetti C,Linnen JM,Kolk DP,Dockter J,Gillotte-Taylor K,Park M,et al. Highly sensitive multiplex assay for detection of human immunodeficiency virus type 1 and hepatitis C virus RNAJ Clin MicrobiolYear: 2002402408241912089255
17. Gallarda JL,Dragon E. Blood screening by nucleic acid amplification technology: current issues, future challengesMol DiagnYear: 20005112210837085
18. Lau LT,Feng XY,Lam TY,Hui HK,Yu AC. Development of multiplex nucleic acid sequence-based amplification for detection of human respiratory tract virusesJ Virol MethodsYear: 20101681-225125420447419
19. Storch GA. Rapid diagnostic tests for influenzaCurr Opin Pediatr.Year: 2003Month: 2 151778412544276
20. Kolk DP,Dockter J,Linnen J,Ho-Sing-Loy M,Gillotte-Taylor K,McDonough SH,et al. Significant closure of the human immunodeficiency virus type 1 and hepatitis C virus preseroconversion detection windows with a transcription-mediated-amplification-driven assayJ Clin MicrobiolYear: 2002401761176611980957
21. Vargo J,Smith K,Knott C,Wang S,Fang C,McDonough S,et al. Clinical specificity and sensitivity of a blood screening assay for detection of HIV-1 and HCV RNATransfusionYear: 20024287688512375660
22. Ayele W,Pollakis G,Abebe A,Fisseha B,Tegbaru B,Tesfaye G,et al. Development of a nucleic acid sequence-based amplification assay that uses gag-based molecular beacons to distinguish between human immunodeficiency virus type 1 subtype C and C' infections in EthiopiaJ Clin MicrobiolYear: 2004421534154115071000
23. Antunes R,Figueiredo S,Bartolo I,Pinheiro M,Rosado L,Soares I,et al. Evaluation of the clinical sensitivities of three viral load assays with plasma samples from a pediatric population predominantly infected with human immunodeficiency virus type 1 subtype G and BG recombinant formsJ Clin MicrobiolYear: 2003413361336712843094

Figures

[Figure ID: F0001]
Fig. 1 

Process of the continuous, homogeneous, isothermal amplification of RNA



[Figure ID: F0002]
Fig. 2 

Comparison between HCV sequences of several genotypes and oligonucleotide sequences. The oligonucleotide sequences were analyzed by Mega4 software for their compatibility.



[Figure ID: F0003]
Fig. 3 

Comparison between HIV-1 sequences of several subtypes. The oligonucleotide sequences were analyzed by Mega4 software for their compatibility.



[Figure ID: F0004]
Fig. 4 

Quality of RNA transcript. Lanes 1, 2 and 3 shows RNA transcript before DNase treatment. Lanes 4, 5 and 6 shows RNA transcript after DNase treatment.



[Figure ID: F0005]
Fig. 5 

Real-time NASBA detection of two different viral genomes in a multiplex format. Each reaction contained two sets of NASBA primers specific for unique HIV-1and HCV nucleotide sequences and two molecular beacons, each specific for one of the two amplicons and labeled with a differently colored fluorophore. HCV is plotted in Solid, Multiplex format plotted in Dotted, HIV is plotted in Dashed and Non template control is plotted in thin line.



[Figure ID: F0006]
Fig. 6 

Shows multiplex NASBA amplification. 241 bp and 179 bp bands in lane1 show simultaneous amplification of HCV and HIV-1, respectively. Lane 2 is a 100 bp RNA marker (Fermentas#SM1831, RNA Ladder 100-1000bp).



Tables
[TableWrap ID: T0001] Table 1 

Sequences primers and probes.


Name Sequence 5’-3’
HIV-1 forward primer GTACAGTGCAGGGGAAAG
HIV-1 Reverse primera AATTCTAATACGACTCACTATAGGG CCAGAGTAGYTTTGCTGGTC
HIV-1 Probeb TET CCCGTGGTYTATTACAGGGACAGCAGAACGGG BHQ-1
HCV forward primer CATGGCGTTAGTAYGAGTG
HCV Reverse primer AATTCTAATACGACTCACTATAGGG CTATCAGGCAGTACCACAAG
HCV Probe FAM CCGATCAGCCATAGTGGTCTGCGGAAGATCGG BHQ-1

aT7 promoter is indicated with Italic

bStem sequences are indicated with underlines


[TableWrap ID: T0002] Table 2 

Limit of quantification Multiplex Real-time NASBA assay for HIV-1a/HCVb detection


aHIV-1 concentration (copies/ml) Number tested Number positive Positive rate (%)
104 10 10 100
103 10 10 100
500 10 3 30
50 10 0 0
bHCV concentration (copies/ml) Number tested Number positive Positive rate (%)
104 10 10 100
103 10 10 100
500 10 10 100
50 10 7 70

aThe LoQ for HIV-1 was determined to be 1000 copies/ml with 95% confidence interval.

bThe LoQ for HCV was determined to be 500 copies/ml with 95% confidence interval.



Article Categories:
  • Original Article

Keywords: HIV-1, HCV, Multiplex NASBA assay, Molecular beacon.

Previous Document:  Redox status is critical for stemness in skin equivalents.
Next Document:  Cloning of fimH and fliC and expression of the fusion protein FimH/FliC from Uropathogenic Escherich...