Document Detail

Microfluidic quadrupole and floating concentration gradient.
Jump to Full Text
MedLine Citation:
PMID:  21897375     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The concept of fluidic multipoles, in analogy to electrostatics, has long been known as a particular class of solutions of the Navier-Stokes equation in potential flows; however, experimental observations of fluidic multipoles and of their characteristics have not been reported yet. Here we present a two-dimensional microfluidic quadrupole and a theoretical analysis consistent with the experimental observations. The microfluidic quadrupole was formed by simultaneously injecting and aspirating fluids from two pairs of opposing apertures in a narrow gap formed between a microfluidic probe and a substrate. A stagnation point was formed at the centre of the microfluidic quadrupole, and its position could be rapidly adjusted hydrodynamically. Following the injection of a solute through one of the poles, a stationary, tunable, and movable-that is, 'floating'-concentration gradient was formed at the stagnation point. Our results lay the foundation for future combined experimental and theoretical exploration of microfluidic planar multipoles including convective-diffusive phenomena.
Authors:
Mohammad A Qasaimeh; Thomas Gervais; David Juncker
Related Documents :
18584665 - Activated sludge systems with biomass particle support structures.
22088045 - Communication: limitations of the stochastic quasi-steady-state approximation in open b...
1366485 - Multimolecular process in a packed-bed immobilized enzyme reactor: numerical simulation...
17530785 - Selective adsorption of volatile organic compounds in micropore aluminum methylphosphon...
14995705 - Noise-induced enhancement of fluctuation and spurious synchronization in uncoupled type...
17155805 - Gain-assisted slow to superluminal group velocity manipulation in nanowaveguides.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2011-09-06
Journal Detail:
Title:  Nature communications     Volume:  2     ISSN:  2041-1723     ISO Abbreviation:  Nat Commun     Publication Date:  2011  
Date Detail:
Created Date:  2011-09-07     Completed Date:  2012-01-18     Revised Date:  2014-04-18    
Medline Journal Info:
Nlm Unique ID:  101528555     Medline TA:  Nat Commun     Country:  England    
Other Details:
Languages:  eng     Pagination:  464     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Microfluidics*
Grant Support
ID/Acronym/Agency:
125907-1//Canadian Institutes of Health Research; 82507-1//Canadian Institutes of Health Research
Comments/Corrections

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-journal-id): 101528555
Journal ID (pubmed-jr-id): 37539
Journal ID (nlm-ta): Nat Commun
Journal ID (iso-abbrev): Nat Commun
ISSN: 2041-1723
Article Information
Download PDF

nihms-submitted publication date: Day: 2 Month: 4 Year: 2014
Electronic publication date: Day: 06 Month: 9 Year: 2011
pmc-release publication date: Day: 11 Month: 4 Year: 2014
Volume: 2First Page: 464 Last Page: 464
PubMed Id: 21897375
ID: 3984239
DOI: 10.1038/ncomms1471
ID: CAMS3944

Microfluidic quadrupole and floating concentration gradient
Mohammad A. Qasaimeh12
Thomas Gervais3
David Juncker124
1Biomedical Engineering Department, McGill University, Montréal, QC H3A 1A4, Canada
2Genome Quebec Innovation Centre, McGill University, Montréal, QC H3A 1A4, Canada
3Department of Engineering Physics, École Polytechnique de Montréal, Montreal, Canada
4Department of Neurology and Neurosurgery, McGill University, Montréal, QC H3A 1A4, Canada
Correspondence: Correspondence and requests for materials should be addressed to D. J. (david.juncker@mcgill.ca)

Introduction

Quadrupoles are used in numerous physical and engineering applications and arise under many forms, such as electrical charges and currents 1, magnetic poles 2, acoustic poles 3,4, and gravitational masses 5. Two popular applications of quadrupoles include the quadrupole mass spectrometer which uses four parallel metal rods with opposing alternating electrical currents to filter ions based on their mass-to-charge ratio 6, and the quadrupole magnets that are used to focus beams of charged particles in particle accelerators 2. In fluid mechanics, flow dipoles, or “doublets” have been studied in porous rocks notably in the context of oil extraction7,8. To increase the recovery of oil pumped from a well, a solution can be injected in another well to displace the oil trapped within the rocks according to a dipolar flow profile. We recently introduced the microfluidic probe (MFP) 9 which uses both injection and aspiration aperture to flush a stream across a substrate surface, and which in fact represents a two-dimensional fluidic dipole, although this analogy has not been developed. Planar multipoles with more than two flow poles have been investigated theoretically by Koplik et al. with increasing numbers of sources and sinks (such as quadrupoles, octopoles, etc.)10. Bazant et al. have also demonstrated the use of conformal mapping to solve a broad class of advection-diffusion problems in 2D irrotational flows11 by drawing an analogy with the pioneering work of Burgers on out-of-plane flow velocity in 2D vortex sheets12. However, experimental validation of these theoretical analyses is still outstanding as flow multipoles have not, to the best of our knowledge, been produced experimentally.

In this work, we present an experimental planar microfluidic quadrupole (MQ) in an open space confined between two parallel plates, Fig. 1. We extend the theoretical framework to describe hydrodynamic and convective-diffusive properties of the MQ in Hele-Shaw flows and derive the equations for the key parameters characteristic of the MQ such as the size of the flow confinement and length of the gradient formed within the MQ. Finite element simulations complete the analytical results and both are compared to experimental results. Finally, we present a first application of the MQ which can be used as a gradient generator and which unlike prior gradient generators affords low shear stress, rapid spatiotemporal tuning of the gradient either hydrodynamically by adjusting the flow rates, or physically by moving the MFP, and which can be applied on any planar substrate and displaced as well.


Results
Microfluidic Quadrupole

MQs were formed between a MFP9 with 4 apertures arranged in a classical symmetric quadrupole configuration (Fig. 1b) and a flat bottom substrate. The MFP was microfabricated into Si and diced into dies with four openings constituting pairs of opposing apertures used for injection and aspiration. A Polydimethylsiloxane (PDMS) adapter block was bonded to the die, and capillaries were plugged into the block and connected to syringe pumps, Fig. 2a,b.

The MFP was immersed in fluid and brought close to a planar substrate so as to form a narrow gap (see the Methods section for more details). A glass slide was used as the bottom substrate permitting visualization of the flow within the gap using an inverted microscope (Fig. 2c). Simultaneous injection and aspiration of fluid generates the quadrupolar flow and the two injected fluids meet head-on at the centre of the MQ, generating a stagnation point (SP) with zero flow velocity (Fig. 1c–f). When the aspiration and injection flow rates are identical, a large fraction of the injected stream is re-aspirated; but the fraction of fluid that radially flows outward is not (see Fig. 1b) and thus leaks into the fluid surrounding the MFP. However, if the ratio of aspiration/injection flow rates is > 1, all the injected liquid is captured and aspirated back into the MFP (Fig. 1f).

We arbitrarily define the two injection apertures as positive poles (sources) and the aspiration apertures as negative poles (drains); see Supplementary Fig. S1 and Supplementary Methods. The MFP used here featured a center-to-center separation of 1075 μm between pairs of opposing apertures, each 360 μm in diameter, and was positioned above, parallel to the bottom substrate so as to form a 50 μm gap for all experiments. This setup represents a parallel-plates configuration, and the MQ formed in the gap is quasi two-dimensional. Using the Hele-Shaw approximation, it can thus be conveniently described as a two-dimensional Stokes flow of the form 13:

[Formula ID: FD1]
(1) 
v⇀H-S(x,y,z)=-H22ηzH(1-zH)∇⇀p(x,y)
where v⇀H-S is the velocity field of the MQ, η is the viscosity, p is the hydrostatic pressure, H is the height of the gap between the plates (H = 50 μm), and z is the vertical coordinate whereas z = 0 corresponds to the bottom plate (substrate) and z = H to the top plate (MFP). The resulting flow velocities in the MQ are on the order of 2 mm/s or less for aspiration flow rates of 100 nL/s and the flow is purely viscous (Re ≪ 1), Supplementary Methods for details. Averaging Equation (1) over the gap height, the velocity profile can be simplified as:
[Formula ID: FD2]
(2) 
v⇀ave(x,y)=1H∫0Hv⇀H-Sdz=-H212η∇⇀p(x,y)

The height-averaged velocity profile given by this expression is irrotational13, and can therefore be represented by a scalar potential akin to an electric field or a current distribution that can also be described based on an electrostatic potential (see Table 1). It thus becomes possible to calculate the flow velocity for any coordinate of the MQ by using the superposition principle as in the case of electrical charges and current quadrupoles (see Supplementary Methods for details):

[Formula ID: FD3]
(3) 
v⇀ave=∑i=14v⇀i=∑i=14Qi(r⇀-r⇀′i)2πH|r⇀-r⇀′i|2
where r⇀ is the position vector and r⇀′i is the position of the sources and drains, both with respect to the center of the probe. Qi is the flow rate from the ith inlet corresponding to Qinj and Qasp for injected and aspirated flow rate, respectively.

Flow confinements of the MQ

As mentioned earlier, the flow rates ratio Qasp/Qinj needs to be sufficiently high to hydrodynamically confine and capture all the injected fluid (See Fig. 1b vs Fig. 1f). The confinement of the injected streams as function of the flow rates ratio were visualized using fluorescent tracer particles, Fig. 3a–c. The area flushed by the injected streams is an important concept for fluidic applications because chemicals injected in the streams may be used to process selectively the underlying surface9, and we thus establish an analytical solution. The fluid that radiates outwards relative to the center of the MQ will travel the greatest distance until the flow velocity becomes zero, at which point the flow direction is reversed and it is eventually recaptured by an aspiration aperture. The confinement radius R of the MQ was defined as the distance between the SP and the outermost point of zero velocity for the radial streams, and using equation (3) yields (see Supplementary Methods for details):

[Formula ID: FD4]
(4) 
R=d2Qasp/Qinj+1Qasp/Qinj-1
where Qasp is the aspiration flow rate, Qinj is the injection flow rate, and d is the center-to-center distance between the two inlets. We found excellent agreement between the calculated and experimentally measured R,Fig. 3. These results confirm the validity of our theoretical assumptions and analysis.

Fluid flow around the SP

A two-variable (x,y) Taylor expansion of equation (3) yields the velocity profile in the very close vicinity of the SP (valid for x,y≪d), see Supplementary Methods for details:

[Formula ID: FD5]
(5) 
v⇀ave=-8〈Q〉πHd2(xx^-yy^)
where 〈Q〉 = (Qasp + Qinj)/2 is the average flow rate between the inlet and outlet, and the origin (0,0) corresponds to the position of the SP. The flow profile can also be expressed in the form of a velocity potential11, see Table 1:
[Formula ID: FD6]
(6) 
ϕ(x,y)=-4〈Q〉πHd2(x2-y2)

Interestingly, this velocity potential is identical to the one describing two-dimensional flow along a right angle corner14. In the MQ, the four quadrants defined by the symmetry lines XX′ and YY′ in Fig. 1d each constitute a virtual corner of a corresponding corner flow. Path lines formed by green tracer beads and a ratio Qasp/Qinj = 10 outline the SP at the center (Fig. 1f and Supplementary Movie 1), and are consistent with the velocity profile and streamlines obtained using FEM simulations, Fig. 4a. As the flow velocity tends to zero when approaching the SP, Fig. 4b, the calculated wall shear stress was found to converge to zero as well, Supplementary Fig. S2a. The shear stress increases linearly when moving away from the SP and reaches a maxima near the inner edge of each aspiration aperture. Within a circle with a diameter of 100 μm (~10% of the distance d) centered at the SP, shear stresses are less than 10% of that local maxima as shown in Fig. 4c (see also Supplementary Methods). For a MQ with injection and aspiration flow rates of 10 nl/s and 100 nl/s respectively, the local maxima of shear stresses is τmax = 0.15 Pa, and hence < 0.015 Pa inside the circle. The shear stress also varies linearly with the aspiration flow rates that may be reduced to further minimize stress (see Supplementary Fig. S2b and Supplementary Methods).

Convection-diffusion and gradients at the SP

When a solute is added to one of the injected streams, diffusive mass transport gradually dominates when approaching the SP as the convective flow converges towards zero. The solute is continuously replenished by one stream acting as a source, while it is also continuously transported away by the other stream which acts as a sink. Along the axis of the two injection apertures (XX′), the solute diffuses into the opposing stream at the SP while it is simultaneously pushed back by the inward convection the opposing stream, and a steady-state, stationary concentration gradient is formed. To characterize this gradient mathematically, we first consider the general time-dependent diffusion-convection equation15:

[Formula ID: FD7]
(7) 
∂C∂t=D∇⇀2C-(∇⇀ϕ)·∇⇀C
where D is the diffusivity of the species studied, v=∇⇀ϕ is the height-averaged in plane velocity profile and C is the concentration of the solute (normalized between 0 and 1). The steady state version of equation (7) (with the left hand term set to zero) is reminiscent of the flow in a “Burgers vortex sheet”12 where C is analogous to the out of plane velocity, as was pointed out by Bazant and Moffatt16.

Based on the experimental observations shown below, the gradient length, which here is defined as the distance between the points with a concentration of 10% and 90% of the original solute concentration, is small when compared to the distance between the inlets along the XX′ axis (see Fig. 5a). Using the approximated velocity profile around the SP of equation (5), and combining it with equation (7), the convection-diffusion equation relevant to the flow around the SP formed within the MQ can be re-written as:

[Formula ID: FD8]
(8) 
∂C(x,t)∂t=D∂2C(x,t)∂x2+8〈Q〉πHd2×∂C(x,t)∂x
where D is the diffusion constant of the solute. The derivatives with respect to y have been neglected in equations (7) and (8) as the concentration interface is invariant in y (∂C/∂y = 0 everywhere) under Hele-Shaw assumption; see Supplementary Methods for details. Using the dimensionless numbers x∼=x/x0=x/(dPé), = t/t0 = · Dt/d2, where Pé=8〈Q〉πHD is the global Péclet number for the MQ, we find the simple dimensionless expression for this transport problem to be:
[Formula ID: FD9]
(9) 
∂C(x∼,t∼)∂t∼=∂2C(x∼,t∼)∂x∼2+x∼∂C(x∼,t∼)∂x∼

Therefore x0=dPé and t0 = d2/(· D) are the natural scales of the MQ. They represent respectively the length over which diffusion takes place at the interface and the characteristic time before steady state is achieved. Typical condition in our experiments were an average flow rate of 55 nl/s, and D = 500 μm2/s as we used fluorescein as the diffusive species. Using these values, we obtain x0 = 15 μm and t0 = 0.4 s. Thus, after a few seconds, the gradient will have reached a steady-state and the time-dependent term of equation (9) can be neglected. When performing this approximation, equation (9) reduces to a one dimensional problem with C(x,t) being the concentration profile at the SP perpendicularly to the interface where diffusion takes place.

For the experimental conditions used here, the concentration gradients at the SP were much shorter than the distance d separating the two inlets was much greater than the gradient length at the SP, d may thus be considered infinite and thus the boundary conditions become:

[Formula ID: FD10]
(10) 
C(∞)=1,C(-∞)=0

This allows simplifying the calculations and solving equations (9) and (10) for the concentration under these conditions gives:

[Formula ID: FD11]
(11) 
C(x)=Pé2π∫-∞x/de-Péξ22dξ=12(1+erf(Pé2xd))
where erf is the error function17. Once again, the solution here is similar to that of the out-of-plane velocity in a Burgers vortex sheet, but with the Péclet number replacing the Reynolds number of the classical analysis16. The gradient length can be calculated using equation (11) and expressed as:
[Formula ID: FD12]
(12) 
Lgrad=2,563d2Pé=1,606·dHD〈Q〉

The gradient length is proportional to the average flow rate and obeys a simple power law.

The analytical solution for the gradient length was compared to experimental measurements using fluorescein sodium salts (376 Da, diffusion coefficient in water D = 500 μm2/s 1820) diluted in water. Fluorescein was injected through the top right aperture, and the gradient formation at the SP within the MQ was modelled numerically (Supplementary Fig. S3a), and measured experimentally, Fig. 5a. Theoretical and experiment gradient profiles are in excellent agreement, Fig. 5b and Supplementary Fig. S3b.

We observed that the gradient was constant along the YY′ axis connecting the two aspiration apertures, Fig. 5a,b. The flow perpendicular to the gradient carries it towards each of the aspiration apertures, and convection rapidly dominates diffusion when moving away from the SP (see Supplementary Methods for details) while at the same time, the concentric flow around the aspiration apertures focuses the stream when approaching the inlet. Collectively, these effects counter diffusive broadening of the gradient which explains the quasi-constant gradient profile along the YY′ line connecting both aspiration apertures.

We compared the calculated gradient length with experimental measurements for a Qasp/Qinj varying from 2 to 15, Fig. 5c–g, and found them to be in good agreement. For Qasp increasing from 20 nl/s to 250 nl/s, the gradient length decreased from 69 μm to 25.7 μm. This range of accessible gradient slopes may be further increased by exploring a wider range of flow rates from pl/s to μl/s. The gradient changes rapidly (see Supplementary Movie 2), and therefore dynamic gradients may readily be formed simply by pre-programming the aspiration flow rates. Whereas the confinement of the MQ is governed by Qasp/Qinj, Equation (12) indicates that the gradient slope and length depend on 〈Q〉 and thus primarily on Qasp for high ratios of Qasp/Qinj. This theoretical result is supported by the experiments where only Qasp was changed, Fig. 5, or where only Qinj was tuned, Supplementary Fig. S4 and Movie 3. It is thus possible to vary the confinement and the gradient slope independently by keeping one of them constant by simultaneously tuning Qasp and Qinj.

Asymmetric MQ flow profiles and Floating Gradient

For all experiments described up to this point, each pair of injection and of aspiration flow rates were adjusted in synchrony and kept equal, which fixed the SP at the center of the MQ. If the flow rate of only one of the two injection or aspiration apertures is changed, the symmetry is broken and the position of SP will change, Fig. 6a,b. The gradient, which originates at the SP moves along with the SP and can thus be displaced hydrodynamically within the central area of the MQ. The gradient is formed along a straight interface when the SP is centered, and along a curved interface when moving the SP toward any of the injection apertures, Fig. 6c,d. By using a programmable flow control system, or by manually changing the flow rates, oscillating, rapidly moving gradients can be generated (see Supplementary Movie 4). Complex spatiotemporal gradient landscapes may thus be formed around a particular point by changing both the gradient slope and the gradient position. The gradient may be adjusted either in a pre-programmed manner, or in real time in response to observations.

Alternatively, for scanning larger areas and reach positions outside the MQ’s central area, the gradient at a point may also be changed by displacing the substrate relative to the MFP, either by moving the bottom substrate which is clamped to the motorized xy-stage of the microscope, or by manually moving the MFP using the micrometer screws of the probe holder (See Fig. 2c) 21. The MQ field lines and the concentration gradient are temporarily perturbed while the MFP moves relative to the substrate proportionally to the speed of displacement. For example, a speed of 300 μm/s lead to a small but visible disturbance of the gradient, Supplementary Movie 5. The results indicate that the gradient re-equilibrated within few seconds after stopping the movement. Faster movements are possible but at the expense of greater disturbance of the gradient. Supplementary Movie 6 shows the gradient while moving the substrate at speed of 7 mm/s, and Supplementary Movie 7 shows the gradient while the MFP is being manually displaced. Using these two approaches, different areas of the substrates can be exposed to the gradient one after the other.


Discussion

We have presented an experimental demonstration of fluidic quadrupoles and found excellent agreement between theoretical prediction and experimental results, thus lending support to the general analytical framework developed to describe multipolar flow. Using the Hele-Shaw formalism, we established a close analogy between MQs, electrical quadrupoles, and vortex sheets in fluid dynamics. Our analysis was extended to include phenomena that were not studied before and the governing equations for a Hele-Shaw geometry derived for flow confinement and for the convection-diffusion of solutes at the SP. In analogy to electrical poles, and as was developed theoretically11,16, the number of poles may be further expanded and inspire combined theoretical and experimental studies of convection-diffusion phenomena with multiple poles and multiple chemicals.

We demonstrated a first application of MQs as concentration gradient generators. The gradient is generated around the SP which experiences no flow and no shear stresses similar to stationary source-sink gradients 22,23, while being continuously fed by the streams of the MQ and thus formed within seconds and could be rapidly adjusted, akin to microfluidic gradients with active flow 2427. Several approaches for forming gradients using hybrid approaches combining flow and no flow zones have been proposed, but required porous elements 28,29 or nanochannels 30,31. Gradients in open chambers have also been proposed 32,33 to allow user access (i.e. pipetting) to the cell culture. These devices provided low shear stress cultures while allowing to adjust the concentration gradients.

However, these devices either entail closed chambers, or complicated fabrication processes, or both, and the time to adjust the gradient was typically longer than with the MQ. Another recent technique34 used a laser beam to open and close pores in an integrated impermeable membrane to expose the cell culture to reagents selectively, however, this device involves closed cell culture chamber and a sophisticated experimental setup.

The floating gradients formed within the MQ are remarkable because of a series of individual characteristics, and because of the properties and possibilities that arise by the combination of these characteristics. They include being open and brought down onto the target areas, the combination of active flow and shear free zone obtained using hydrodynamic effects alone and without need for nanochannels or membranes, the possibility to rapidly adjust and move the gradient hydrodynamically within a few seconds, while also allowing to move it physically by displacing the MFP, and finally the possibility of repeated use of the device for many different experiments

This setup should be well suited for cell biological applications such as studying cellular migrations28 and neuronal navigation3537, or stem cell differentiation38 because chemical cues can be applied and modulated rapidly to individual cells at the SP with minimal shear stress. It will be interesting to find out whether these properties can be exploited and the MFP-MQ used to study the response of neurons and stem cells to gradients; indeed neurons and stem cells are particularly difficult to culture and maintain inside closed channels for long periods39,40, and that have been shown to be very sensitive to shear stress41. Furthermore, the MQ and floating gradients could be used to study large samples which are hard to cultivate inside closed channels, such as embryos or tissue slices42. Finally, it may be useful for surface patterning by positioning and scanning the MFP atop surfaces9,43,44.


Methods
The MFP Fabrication and Operation

The MFP comprises a square Si chip (3 mm in width) with four etched holes of 360 μm diameter, and a PDMS interface chip with holes for receiving the capillaries of 360 μm and 250 μm outer and inner diameters, respectively, that connect it to the injection and aspiration syringe pumps (Nemesys, Cetoni, Germany ) see Fig. 2a. The Si chips were fabricated in 200 μm thick, double-side-polished silicon wafers (Silicon Quest International Inc, California), using photolithography and deep-reactive-ion etching. Etching masks were formed on silicon wafers by using standard photolithography and S1813 photoresist (Shipley, Massachusetts) to pattern a 2 μm thick thermal SiO2 layer. The openings were made by deep reactive ion etching (DRIE, ASE etcher, Surface Technology Systems). The PDMS interface block was fabricated by casting poly(dimethylsiloxane) PDMS (Sylgard 184, Dow Corning, Midland, Michigan) into a home-made micromold composed of two structured steel plates, a polished steel plate with vias (access holes) forming the bottom, and four capillaries (each inserted into one of the four vias in the polished steel plate) serving as place holders for the fluidic connection holes. The PDMS was cured in an oven at 60 °C for at least 3 h (usually over night). The PDMS block was bonded to the Si chip by activating both parts in air plasma (Plasmaline 415 Plasma Asher, Tegal Corporation, California), at 0.2 mbar for 45 s at 75 W, and joining the two together using a home-made mechanical alignment aid and placed in an oven at 90 °C for 20 minutes. Operation of this MFP is similar to the use of the original MFP 8,9,21, except that four independent syringe pumps are required to control the flow rate of each aperture independently. Briefly, the MFP was secured inside the clamping rode (Fig. 2b) and mounted on the xyz micropositioner (7600-XYZL, Siskiyou inc, Oregon), what we call the probe holder. Using the probe holder, the MFP was positioned parallel to the transparent substrate (glass slide) atop an inverted microscope (TE2000, Nikon) so as to form a microscopic gap while immersed under the surrounding medium, Fig. 2c. Capillaries connect each aperture is glass syringes (Hamilton, Reno, NV) that are operated using a computer-controlled syringe pumps (Nemesys Cetoni, Germany) through and capillaries (Polymicro Tech, Arizona).

Numerical Simulations

The 3D simulations were carried out using the commercially available finite element simulation software Comsol Multiphysics 3.5 (Comsol Inc, Massachusetts) and run on an eight-core, 64 bit computer (Xeon, Dell) with 26 GB of RAM. The MFP geometry was modeled with the same shape and dimensions as the microfabricated MFP used in the experiments. Simulations coupled the solution of Navier-Stokes equation and convection-diffusion equation. Sample solutions were assumed to be water and incompressible with a density of 998.2 kg/m3, and a dynamic viscosity of 0.001 N·s/m2), and the diffusion coefficient of the solute was 500 μm2/s, which corresponds to the diffusion constant of fluorescein in water. The simulations were run under steady state conditions and assumed no-slip boundary conditions on the substrate and the MFP surface, with the flow boundary conditions at the MQ’s perimeter sides set as open boundaries (equal to atmospheric pressure). As boundary conditions for the mass transfer equation, the MFP surface and substrate were defined as an insulating boundary, and a zero inward flux was set at the perimeter of the MQ model. The experimental values were used for injection and aspiration flow rates. The concentration of the solute at the source injection aperture was arbitrarily set to 1 mol/m3, 0 mol/m3 at the other injection aperture, and convective flux was set for both aspiration apertures.

Image Acquisition and Analysis

Images were recorded using a cooled CCD camera (Photometrics CoolSNAP HQ2) connected to the microscope. Black-and-white images were analysed and coloured using the software ImageJ 1.42 (National Institutes of Health Maryland). Images of green and red beads pathlines were joined together digitally using the freeware software GIMP 2.6.7 (Free Software Foundation, Inc., Massachusetts). Videos were recorded using a digital consumer camcorder camera (HDR-SR7, Sony Electronics Inc, California) connected to the microscope through a C-mount adaptor. The gradient lengths shown in Fig. 5g were defined as the distance between 90 percent and 10 percent of the maximum fluorescence intensity signal, as shown in Supplementary Fig. S5.

Chemicals and Experimental Procedures

Fluorescein sodium salt (C20H10Na2O5) was purchased from Sigma-Aldrich (Sigma-Aldrich Canada Ltd, Ontario) and fluorescein solutions were prepared by dissolving 0.03 g of the salt with 100 ml of distilled water. Red and yellow-green carboxylate-modified fluorescent microspheres (2 μm diameters) were purchased from Invitrogen (fluosphere, Invitrogen Canada Inc, Ontario). The solutions used in experiments were prepared by diluting 0.1 ml of the beads solution with 13 ml of distilled water.



Notes

FN2Author contributions

M.A.Q. designed the research, carried out the FEM analysis, performed the experiments, analysed the results, and wrote the article; T.G. carried out the mathematical analysis and wrote the article; D.J. designed the research and wrote the article.

FN3Additional information

The authors declare no competing financial interests.

The authors acknowledge funding from NSERC, CIHR, CHRP, Genome Canada, Genome Quebec, CFI, and the assistance of McGill Nanotools Microfab Laboratory (funded by CFI, NSERC and VRQ). M.A.Q. thanks Roozbeh Safavieh and Cécile Perrault for their help and discussion. We acknowledge Michel Godin, Sebastien Ricoult, Gina Zhou, Huiyan Li, Setareh Ghorbanian, and Nageswara R. Ghattamaneni for critical reading the manuscript. M.A.Q. acknowledges Alexander Graham Bell Canada NSERC Scholarship and D.J. acknowledges support from a Canada Research Chair.


References
1. Jackson JD. Classical Electrodynamics3832John Wiley & SonsYear: 1999
2. Wiedemann H. Particle Accelerator Physics. I. Basic Principles and Linear Beam Dynamics2469SpringerYear: 1999
3. Russell DA,Titlow JP,Bemmen YJ. Acoustic monopoles, dipoles, and quadrupoles: An experiment revisitedAm J Phys67660664Year: 1999
4. Meyer Erwin,Neumann E-G. Physical and Applied Acoustics: An Introduction412Academic PressYear: 1972
5. Thorne KS. Multipole expansions of gravitational radiationRev Mod Phys52299Year: 1980
6. de Hoffmann Edmond,Stroobant V. Mass Spectrometry: Principles and Applications3503John Wiley & SonsYear: 2007
7. Scheidegger AE. The physics of flow through porous mediaUniversity of Toronto PressYear: 1974
8. Bear J. Dynamics of Fluids in Porous MediaAmerican ElsevierYear: 1972
9. Juncker D,Schmid H,Delamarche E. Multipurpose microfluidic probeNat Mater4622628Year: 200516041377
10. Koplik J,Redner S,Hinch EJ. Tracer dispersion in planar multipole flowsPhys Rev E504650Year: 1994
11. Bazant MZ. Conformal Mapping of Some Non-Harmonic Functions in Transport TheoryProc R Soc A46014331452Year: 2004
12. Burgers JM. Richard Von Mises,Theodore Von KármánAdv Appl Mech1171199ElsevierYear: 1948
13. Batchelor GK. An Introduction to Fluid DynamicsCambridge University PressYear: 1967
14. Batchelor GK. An Introduction to Fluid DynamicsCambridge University PressYear: 2000
15. Deen WM. Analysis of Transport Phenomena624Oxford University PressYear: 1998
16. Bazant MZ,Moffatt HK. Exact solutions of the Navier Stokes equations having steady vortex structuresJ Fluid Mech5415564Year: 2005
17. Abramowitz MA,Stegun IA. Handbook of Mathematical Functions: with Formulas, Graphs, and Mathematical Tables1Dover seriesYear: 1964
18. deBeer D,Stoodley P,Lewandowski Z. Measurement of local diffusion coefficients in biofilms by microinjection and confocal microscopyBiotechnol Bioeng53151158Year: 199718633959
19. Wang LY,et al. In situ measurement of solute transport in the bone lacunar-canalicular systemProc Natl Acad Sci U S A1021191111916Year: 200516087872
20. Rani SA,Pitts B,Stewart PS. Rapid diffusion of fluorescent tracers into Staphylococcus epidermidis biofilms visualized by time lapse microscopyAntimicrob Agents Ch49728732Year: 2005
21. Perrault CM,Qasaimeh MA,Juncker D. The Microfluidic Probe: Operation and Use for Localized Surface ProcessingJ Vis Expe1418Year: 2009
22. Abhyankar VV,Lokuta MA,Huttenlocher A,Beebe DJ. Characterization of a membrane-based gradient generator for use in cell-signaling studiesLab Chip6389393Year: 200616511622
23. Frevert CW,Boggy G,Keenan TM,Folch A. Measurement of cell migration in response to an evolving radial chemokine gradient triggered by a microvalveLab Chip6849856Year: 200616804588
24. Keenan TM,Folch A. Biomolecular gradients in cell culture systemsLab Chip83457Year: 200818094760
25. Toetsch S,Olwell P,Prina-Mello A,Volkov Y. The evolution of chemotaxis assays from static models to physiologically relevant platformsIntegr Biol1170181Year: 2009
26. Jeon NL,et al. Neutrophil chemotaxis in linear and complex gradients of interleukin-8 formed in a microfabricated deviceNat Biotechnol20826830Year: 200212091913
27. Barkefors I,et al. Endothelial cell migration in stable gradients of vascular endothelial growth factor a and fibroblast growth factor 2 - Effects on chemotaxis and chemokinesisJ Biol Chem2831390513912Year: 200818347025
28. Kim T,Pinelis M,Maharbiz M. Generating steep, shear-free gradients of small molecules for cell cultureBiomed Microdevices11657310.1007/s10544-008-9210-7Year: 200918688724
29. Diao J,et al. A three-channel microfluidic device for generating static linear gradients and its application to the quantitative analysis of bacterial chemotaxisLab Chip6381388Year: 200616511621
30. Paliwal S,et al. MAPK-mediated bimodal gene expression and adaptive gradient sensing in yeastNature4464651Year: 200717310144
31. Shamloo A,Ma N,Poo M-m,Sohn LL,Heilshorn SC. Endothelial cell polarization and chemotaxis in a microfluidic deviceLab Chip812921299Year: 200818651071
32. Keenan TM,Hsu CH,Folch A. Microfluidic “jets” for generating steady-state gradients of soluble molecules on open surfacesAppl Phys Lett89114103Year: 2006
33. Cate DM,Sip CG,Folch A. A microfluidic platform for generation of sharp gradients in open-access cultureBiomicrofluidics404410504411010.1063/1.3490784Year: 2010
34. Moorjani S,Nielson R,Chang XA,Shear JB. Dynamic remodeling of subcellular chemical gradients using a multi-directional flow deviceLab Chip1021392146Year: 201020544072
35. Cosson S,Kobel SA,Lutolf MP. Capturing Complex Protein Gradients on Biomimetic Hydrogels for Cell-Based AssaysAdv Funct Mater1934113419Year: 2009
36. Dertinger SKW,Jiang XY,Li ZY,Murthy VN,Whitesides GM. Gradients of substrate-bound laminin orient axonal specification of neuronsProc Natl Acad Sci U S A991254212547Year: 200212237407
37. Yam PT,Langlois SD,Morin S,Charron F. Sonic Hedgehog Guides Axons through a Noncanonical, Src-Family-Kinase-Dependent Signaling PathwayNeuron62349362Year: 200919447091
38. van Noort D,et al. Stem cells in microfluidicsBiotechnol Prog255260Year: 200919205022
39. Regehr KJ,et al. Biological implications of polydimethylsiloxane-based microfluidic cell cultureLab Chip9Year: 2009
40. Millet LJ,Stewart ME,Sweedler JV,Nuzzo RG,Gillette MU. Microfluidic devices for culturing primary mammalian neurons at low densitiesLab Chip7987994Year: 200717653340
41. Kim IA,et al. Effects of mechanical stimuli and microfiber-based substrate on neurite outgrowth and guidanceJ Biosci Bioeng101120126Year: 200616569606
42. Queval A,et al. Chamber and microfluidic probe for microperfusion of organotypic brain slicesLab Chip10326334Year: 201020091004
43. Lovchik RD,Drechsler U,Delamarche E. Multilayered microfluidic probe headsJ Micromech Microeng115006Year: 2009
44. Perrault CM,et al. Integrated microfluidic probe stationRev Sci Instrum81115107Year: 201021133501