Measurement of pulmonary CO2 elimination must exclude inspired CO2 measured at the capnometer sampling site.
MedLine Citation:
PMID:  8823647     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: The pulmonary elimination of the volume of CO2 per breath (VCO2/br, integration of product of airway flow (V) and PCO2 over a single breath) is a sensitive monitor of cardio-pulmonary function and tissue metabolism. Negligible inspired PCO2 results when the capnometry sampling site (SS) is positioned at the entry of the inspiratory limb to the airway circuit. In this study, we test the hypothesis that moving SS lungward will result in significant inspired CO2 (VCO2[I]), that needs to be excluded from VCO2/br. METHODS: We ventilated a mechanical lung simulator with tidal volume (VT) of 800 mL at 10 breaths/min. CO2 production, generated by burning butane in a separate chamber, was delivered to the lung. Airway V and PCO2 were measured (Capnomac Ultima, Datex), digitized (100 Hz for 60 s), and stored by microcomputer. Then, computer algorithms corrected for phase differences between V and PCO2 and calculated expired and inspired VCO2 (VCO2[E] and VCO2[I]) for each breath, whose difference equalled overall VCO2/br. The lung and Y-adapter (where the inspiratory and expiratory limbs of the circuit joined) were connected by the SS and a connecting tube in varying order. RESULTS: During ventilation of the lung model (VT = 800 ml) with SS adjacent to the inspiratory limb, VCO2[E] was 16.8 +/- 0.4 ml and VCO2[I] was 1.1 +/- 0.1 ml, resulting in overall VCO2/br (VCO2[E] - VCO2[I]) of 15.7 +/- 0.4 ml. If VCO2[I] was ignored in the determination of VCO2/br, then the %error that VCO2[E] overestimated VCO2/br was 7.2 +/- 0.3%. This %error significantly increased (p < 0.05, Student's t-test) when VT was decreased to 500 mL (%error = 12.4 +/- 0.8%) or when SS was moved to the lungward side of a 60 mL connecting tube (VCO2[I] = 2.8 +/- 0.2, %error = 18.2 +/- 1.6) or a 140 mL tube (VCO2[I] = 5.9 +/- 0.3 mL, %error = 37.5 +/- 3.3). CONCLUSIONS: When the SS was moved lungward from the inspiratory limb, instrumental dead space (VDINSTR) increased and, at end-expiration, contained exhaled CO2 from the previous breath. During the next inspiration, this CO2 was rebreathed relative to SS (i.e. VCO2[I]), and contributed to VCO2[E]. Thus, VCO2[E] overestimated VCO2/br (%error) by the amount of rebreathing, which was exacerbated by larger VDINSTR (increased VCO2[I]) or smaller VT (increased VCO2[I]-to-VCO2/br ratio).
Authors:
P H Breen; E R Serina; S J Barker
Related Documents :
9244017 - Gastric tonometry and prediction of outcome in the critically ill. arterial to intramuc...
1521047 - Acute ventilation-perfusion mismatching resulting from inhalative smoking of the first ...
11337737 - Abdominal wall lift systems in laparoscopic surgery: gasless and low-pressure systems.
3564877 - Rebreathing and ventilatory response to different fresh gas flows in the bain and lack ...
1495167 - Ventricular contractility evaluated by mechanical perturbation of the myocardium--exper...
17701537 - Monitoring of whole-body hyperthermia with transesophageal echocardiography (tee).
Publication Detail:
Type:  Journal Article; Research Support, U.S. Gov't, P.H.S.    
Journal Detail:
Title:  Journal of clinical monitoring     Volume:  12     ISSN:  0748-1977     ISO Abbreviation:  J Clin Monit     Publication Date:  1996 May 
Date Detail:
Created Date:  1996-12-09     Completed Date:  1996-12-09     Revised Date:  2007-11-14    
Medline Journal Info:
Nlm Unique ID:  8506707     Medline TA:  J Clin Monit     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  231-6     Citation Subset:  IM    
Affiliation:
Department of Anesthesiology, University of California at Irvine Medical Center, Orange 92613, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Capnography / instrumentation,  methods*
Humans
Lung / physiology
Models, Structural
Pulmonary Ventilation
Grant Support
ID/Acronym/Agency:
HL-42637/HL/NHLBI NIH HHS

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine