Mathematical model for the calculation of oxygen concentrations in a closed circuit oxygen rebreathing apparatus.
MedLine Citation:
PMID:  9262815     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Closed circuit oxygen rebreathing diving apparatus are used by armed forces in special tasks because of their advantages of long endurance, low noise and minimal gas escape. There is little knowledge about the administered oxygen concentrations in these systems. Closed circuit oxygen rebreathing apparatus are also used as a first aid device for the treatment of severe disorders. Because of similar constructive components, these rebreathing apparatus are comparable to the Dräger LAR V model. HYPOTHESIS: This study was conducted to measure the oxygen concentrations in the LAR V and estimate the correlation between oxygen concentration and pre-breathing purges. METHOD: Subjects were 12 males who performed the pre-breathing procedure. The oxygen concentrations in the breathing loop were measured after each purge. RESULTS: The oxygen concentrations depended on the volume of the apparatus dead space, the total capacity of the divers breathing system respective to the volume of the breathing purges and the number of pre-breathing purges. The maximum oxygen concentration was reached after eight purges (O2 = 85%). An equation to estimate the oxygen concentration inside the LAR V and first aid rebreather was derived. CONCLUSIONS: The results indicated that the present purging procedure (three purges) before diving is normally not sufficient to remove the nitrogen totally from the deadspace of the LAR V, the divers airways and lungs. Only a small modification (six purges) is necessary to improve the safety of the diver in case of a diving apparatus malfunction during the mission profile.
Authors:
B Neubauer; R Zander; K Tetzlaff
Related Documents :
986615 - Changes in respiratory frequency and end-expiratory volume accompanying augmented breat...
468635 - Breath-by-breath variation of frc: effect on vo2 and vco2 measured at the mouth.
3948635 - Phenyl isopropyl adenosine alters fetal electrocorticogram, breathing and heart rate.
15878235 - Dynamic stability of locomotor respiratory coupling during cycling in humans.
9580855 - Impact of exercise rehabilitation on cardiac neuronal function in heart failure: an iod...
1173445 - Osteoarthritis in knees of aged persons. relationship between roentgenographic and clin...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Aviation, space, and environmental medicine     Volume:  68     ISSN:  0095-6562     ISO Abbreviation:  Aviat Space Environ Med     Publication Date:  1997 Aug 
Date Detail:
Created Date:  1997-10-02     Completed Date:  1997-10-02     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7501714     Medline TA:  Aviat Space Environ Med     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  722-5     Citation Subset:  IM; S    
Affiliation:
Federal German Navy Weapons Diver Battalion, Eckernförde, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Administration, Inhalation
Adult
Breath Tests
Diving* / adverse effects
Equipment Design
Humans
Linear Models*
Male
Mathematics*
Military Personnel*
Oxygen / administration & dosage*,  analysis*
Predictive Value of Tests
Pulmonary Gas Exchange*
Reproducibility of Results
Respiration*
Vital Capacity
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Arm arterial occlusion cuffs as a means of alleviating high +Gz-associated arm pain.
Next Document:  Changes in radioactive tracer distribution in rats after 24 hours of 45 degrees hind limb unweightin...