Magnetohydrodynamic interference with the edge pedestal motional Stark effect diagnostic on DIII-D.
MedLine Citation:
PMID:  21456744     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
Accurate measurement of internal magnetic field direction using motional Stark effect (MSE) polarimetry in the edge pedestal is desired for nearly all tokamak scenario work. A newly installed 500 kHz 32-channel digitizer on the MSE diagnostic of DIII-D allows full spectral information of the polarimeter signal to be recovered for the first time. Fourier analysis of this data has revealed magnetohydrodynamic (MHD) fluctuations in the plasma edge pedestal at ρ ≥ 0.92. By correlating edge localized mode fluctuations seen on lock-in amplifier outputs with MSE spectrograms, it has been shown that edge pedestal tearing mode fluctuations cause interference with MSE second harmonic instrument frequencies. This interference results in unrecoverable errors in the real-time polarization angle measurement that are more than an order of magnitude larger than typical polarimeter uncertainties. These errors can cause as much as a 38% difference in local q. By using a redundant measure of the linear polarization found at the fourth harmonic photo-elastic modulator (PEM) frequency, MHD interference can be avoided. However, because of poorer signal-to-noise the fourth harmonic signal computed polarization angle shows no improvement over the MHD polluted second harmonics. MHD interference could be avoided in future edge pedestal tokamak polarimeters by utilizing PEMs with higher fundamental frequencies and a greater separation between their frequencies.
Authors:
J D King; M A Makowski; C T Holcomb; S L Allen; D N Hill; R J La Haye; F Turco; C C Petty; M A Van Zeeland; T L Rhodes; W H Meyer; R Geer; E C Morse
Related Documents :
8865634 - Psychometric functions for the discrimination of spectral variance.
21047294 - Evaluation of two algorithms for detecting human frequency-following responses to voice...
15553654 - Avoiding electromagnetic artifacts when recording auditory steady-state responses.
20456914 - Pitch-matching accuracy in trained singers and untrained individuals: the impact of mus...
3999044 - The contrast sensitivity of human colour vision to red-green and blue-yellow chromatic ...
15950374 - Exposure to light-at-night increases the growth of dmba-induced mammary adenocarcinomas...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  The Review of scientific instruments     Volume:  82     ISSN:  1089-7623     ISO Abbreviation:  Rev Sci Instrum     Publication Date:  2011 Mar 
Date Detail:
Created Date:  2011-04-04     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0405571     Medline TA:  Rev Sci Instrum     Country:  United States    
Other Details:
Languages:  eng     Pagination:  033515     Citation Subset:  IM    
Affiliation:
Lawrence Livermore National Laboratory, Livermore, California 94550-9234, USAUniversity of California-Berkeley, Berkeley, California 94720-7300, USAGeneral Atomics, P.O. Box 85608, San Diego, California 92186-5608, USAOak Ridge Institute for Science Education, Oak Ridge, Tennessee 37830-8050, USAUniversity of California-Los Angeles, PO Box 957099, Los Angeles, California 90095-7099, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Low noise K(?)-band hopping reflectometer based on yttrium iron garnet sources at TEXTOR.
Next Document:  A combined ion-sputtering and electron-beam annealing device for the in vacuo postpreparation of sca...