MRI of the brain is safe in patients implanted with the vagus nerve stimulator.
MedLine Citation:
PMID:  11749109     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Metallic devices generally represent a contra-indication for MRI scanning. Based on laboratory testing, the neuro cybernetic prosthesis (NCP) is labelled MRI compatible when used with a send and receive head coil. However, there are no published clinical data to support the safety of brain MRI in patients with the NCP. Our objective was to report clinical experience with such a population. We questioned 40 centres that had implanted the NCP system as of 10/1/99. If MRI had been performed on any vagus nerve stimulator patients, we collected information on these patients, the MRI technique used, any events noted during the scan, including both subjective reports (by the patient ), and observable (objective) changes noted by the staff. Twelve centres (30%) responded. Over a time period of 3 years, there were a total of 27 MRI scans performed in 25 patients. All scanners were 1.5 T. A head coil was used in 26 scans, and a body coil in one. The indications for the scans were diverse. Seven were related to the epilepsy, including aetiology or pre-surgical evaluation. Others were unrelated, including brain tumours, cerebral haematoma, vasculitis, headaches, and head trauma. Three scans were performed with the stimulator on, while 24 were performed with the stimulator off. One patient had a mild objective voice change for several minutes. No other objective changes were noted in any of the patients. One 11-year old reported chest pain while experiencing severe claustrophobia. Twenty-five patients denied any discomfort around the lead or the generator. We conclude that this clinical series supports the safety of routine brain MRI using a send and receive head coil in patients implanted with the NCP System.
Authors:
S R Benbadis; J Nyhenhuis; W O Tatum; F R Murtagh; M Gieron; F L Vale
Related Documents :
8829559 - Brain relaxation and cerebrospinal fluid pressure during craniotomy for resection of su...
9151009 - Neurobiological findings in first admission patients with schizophrenia or schizophreni...
19624389 - Microstructural white matter changes in euthymic bipolar patients: a whole-brain diffus...
6127989 - Somatostatin-like immunoreactivity in the cerebrospinal fluid of neurological patients.
8856789 - Severe stealth virus encephalopathy following chronic-fatigue-syndrome-like illness: cl...
11255629 - Brain hypothermia relieves severe brain swelling following acute major cerebral artery ...
10421039 - Home intravenous antibiotic therapy for patients with infective endocarditis.
23717059 - Korean red ginseng powder in the treatment of melasma: an uncontrolled observational st...
19668399 - False negative results in glaucoma detection with heidelberg retina tomograph ii.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Seizure     Volume:  10     ISSN:  1059-1311     ISO Abbreviation:  Seizure     Publication Date:  2001 Oct 
Date Detail:
Created Date:  2001-12-25     Completed Date:  2002-02-28     Revised Date:  2014-10-21    
Medline Journal Info:
Nlm Unique ID:  9306979     Medline TA:  Seizure     Country:  England    
Other Details:
Languages:  eng     Pagination:  512-5     Citation Subset:  IM    
Copyright Information:
Copyright 2001 BEA Trading Ltd.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Brain / anatomy & histology*
Child
Child, Preschool
Electric Stimulation / instrumentation
Electrodes, Implanted
Equipment Safety
Female
Humans
Infant
Magnetic Resonance Imaging*
Male
Middle Aged
Vagus Nerve / physiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Reversed visual field constrictions in children after vigabatrin withdrawal--true retinal recovery o...