Abstract/OtherAbstract:
Experiments were conducted in which various intravascular filters, stents, and coils were imaged using magnetic resonance (MR) spin-echo technique at 0.35 T. These devices were also evaluated for ferromagnetism (at 0.35, 1.5, and 4.7 T), magnetic torque (at 0.35 and 1.5 T), and magnetically induced migration within a plastic tube (at 0.35 and 1.5 T for the Greenfield filter [GF]). The stainless-steel GF was evaluated in vitro for its propensity to perforate canine inferior venae cavae (IVC). Magnetic force and torque at 1.5 T did not dislodge the GF or result in perforation of canine IVC by the GF. Beta-3 titanium alloy (used in a new percutaneous version of the GF) is apparently one of the best-suited metals for use with MR imaging because of its lack of ferromagnetism (up to 4.7 T) and absence of MR imaging artifacts (at 0.35 T). Devices composed of Elgiloy (Mobin-Uddin filter), nitinol, and MP32-N (Amplatz filter) alloys all created mild artifacts. Devices fashioned from 304 and 316L (GF and Palmaz stent) stainless-steel alloys created severe "black-hole" artifacts, with the 304 alloy devices also showing marked image distortion. Generally, the greater the ferromagnetism of a device, the greater its magnetic susceptibility artifact.
Authors:
G P Teitelbaum; W G Bradley; B D Klein
Related Documents :
10805129 - Sensitivity enhancement in (hca)conh experiments.
7811609 - Magnetic resonance angiography.
8387549 - Overview of mr imaging modalities.
21489629 - Mri versus ct for the diagnosis of lumbar spinal stenosis.
7476089 - The application of magnetization transfer to mr angiography with reduced total power.
9090599 - Peripheral nerve stimulation in a whole-body echo-planar imaging system.
19769449 - Multiple sclerosis lesions: insights from imaging techniques.
11417619 - Acute necrotizing encephalopathy presenting as a basal ganglia syndrome.
17318549 - Diagnostic accuracy of fdg pet in the follow-up of platinum-sensitive epithelial ovaria...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Radiology     Volume:  166     ISSN:  0033-8419     ISO Abbreviation:  Radiology     Publication Date:  1988 Mar 
Date Detail:
Created Date:  1988-03-16     Completed Date:  1988-03-16     Revised Date:  2001-03-23    
Medline Journal Info:
Nlm Unique ID:  0401260     Medline TA:  Radiology     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  657-64     Citation Subset:  AIM; IM    
Affiliation:
Department of Radiology, Thomas Jefferson University, Philadelphia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Filtration / instrumentation
Magnetic Resonance Imaging*
Magnetics

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Aortic dissection: sensitivity and specificity of MR imaging.
Next Document:  Common bile duct calculi: updated experience with dissolution with methyl tertiary butyl ether.