Longevity of implantable cardioverter-defibrillators, influencing factors, and comparison to industry-projected longevity.
MedLine Citation:
PMID:  19959121     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Because the best possible device longevity is crucial (i.e., risk of infection with premature device exchange, current cost-effectiveness calculations depending on reasonable longevity, patient comfort), industry-independent real-life data are fundamental. However, only limited independent data on the longevity of implantable cardioverter-defibrillators (ICDs) are available.
OBJECTIVE: The purpose of this study was to determine ICD device longevity and influencing factors.
METHODS: From a prospective database, we studied overall device longevity and identified those devices with replacement for battery depletion or prolonged charge time. For every device, we determined factors that included averaged shocks, pacing percentage, pacing mode, device size, and time of implant. Survival probabilities at different time intervals were calculated, and Kaplan-Meier and Cox regression analyses were used. Observed longevity was compared to industry-projected longevity obtained from product performance reports.
RESULTS: A total of 644 ICDs (Medtronic 317, Guidant 189, St. Jude 118, Intermedics 20) were implanted in 499 patients. During follow-up, 163 (25.3%) ICDs were replaced. Manufacturer, time of implant, pacing mode, pacing percentage, and capacitor reformation interval influenced longevity, whereas device size and number of shocks did not. Median longevity was 7.6 years for Medtronic devices, 5.0 years for Guidant devices, and 3.8 years for St. Jude devices. After 5 years, only 70% of ICDs were still in service compared to the 80% projected by industry.
CONCLUSION: Marked differences in device longevity among manufacturers cannot be explained by pacing mode, number of shocks, or pacing percentage only. Overall, device performance requires further improvement for the sake of patient health and cost.
Authors:
Beat Andreas Schaer; Michael Theodor Koller; Christian Sticherling; David Altmann; Lucas Joerg; Stefan Osswald
Publication Detail:
Type:  Comparative Study; Journal Article     Date:  2009-09-11
Journal Detail:
Title:  Heart rhythm : the official journal of the Heart Rhythm Society     Volume:  6     ISSN:  1556-3871     ISO Abbreviation:  Heart Rhythm     Publication Date:  2009 Dec 
Date Detail:
Created Date:  2009-12-04     Completed Date:  2011-01-28     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101200317     Medline TA:  Heart Rhythm     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1737-43     Citation Subset:  IM    
Affiliation:
Department of Cardiology, University of Basel Hospital, Basel, Switzerland. bschaer@uhbs.ch
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Confidence Intervals
Death, Sudden, Cardiac / prevention & control
Defibrillators, Implantable* / classification,  standards
Equipment Design
Equipment Failure
Female
Heart Failure / therapy
Humans
Kaplan-Meier Estimate
Male
Middle Aged
Multivariate Analysis
Proportional Hazards Models
Risk Factors
Time Factors
Comments/Corrections
Comment In:
Heart Rhythm. 2009 Dec;6(12):1744   [PMID:  19959122 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Evidence of clinical practice heterogeneity in the use of implantable cardioverter-defibrillators in...
Next Document:  Association of TGFBR2 polymorphism with risk of sudden cardiac arrest in patients with coronary arte...