Document Detail


Long term survival of an hydroxyapatite-coated threaded cup in the presence of a high polythene wear rate.
MedLine Citation:
PMID:  20640996     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
We describe the long term clinical results and polythene wear rate measurement of 144 uncemented total hip arthroplasties in 118 patients (Male: Female-65: 53, Mean age: 52.8 years (range 21-78 years) performed between 1988 and 2000 using the Furlong HAC coated threaded acetabular cup. The mean follow-up for the group was 10.2 years (range: 5-17.5, median: 9.7). One femoral stem and two acetabular shells were revised due to aseptic loosening. The mean polythene wear rate was 0.24 mm/year. Ten-year survival for the acetabular and femoral components with radiological evidence of aseptic loosening as an end point was 99.15 (CI: 98.3-99.9) and 99.28 (CI: 98.5-99.9). There was no evidence of osteolysis around the femoral or acetabular components in spite of a relatively high polythene wear rate (0.24 mm/year). Our study demonstrates excellent survival of threaded HAC coated acetabular sockets at 10 years in spite of a relatively high polythene wear rate.
Authors:
Sandeep P Datir; Peter D Angus
Related Documents :
8666626 - The frequency of venous thrombosis in cemented and non-cemented hip arthroplasty.
11476306 - Total hip arthroplasty using porous-coated femoral components in patients with rheumato...
20417906 - The charnley stem: clinical, radiological and survival data after 11-14 years.
9590636 - Total hip arthroplasty in juvenile chronic arthritis: a consecutive series.
17476596 - Remifentanil compared with midazolam and pethidine sedation during colonoscopy: a prosp...
22561216 - Patient-reported long-term rectal function after permanent interstitial brachytherapy f...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Hip international : the journal of clinical and experimental research on hip pathology and therapy     Volume:  20     ISSN:  1724-6067     ISO Abbreviation:  Hip Int     Publication Date:    2010 Jul-Sep
Date Detail:
Created Date:  2010-07-26     Completed Date:  2010-10-28     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9200413     Medline TA:  Hip Int     Country:  Italy    
Other Details:
Languages:  eng     Pagination:  327-34     Citation Subset:  IM    
Affiliation:
Department of Orthopaedics and Trauma, Dewsbury and District Hospital, Dewsbury, West Yorks, UK. datirsandeep@hotmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Arthroplasty, Replacement, Hip / adverse effects,  instrumentation*,  methods
Coated Materials, Biocompatible*
Durapatite
Female
Hip Joint / pathology,  radiography,  surgery*
Hip Prosthesis*
Humans
Male
Middle Aged
Polyethylene
Prosthesis Design
Prosthesis Failure*
Time Factors
Young Adult
Chemical
Reg. No./Substance:
0/Coated Materials, Biocompatible; 1306-06-5/Durapatite; 9002-88-4/Polyethylene

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Retroperitoneal abscess following infected bipolar hemiarthroplasty diagnosed by metallosis: a case ...
Next Document:  Use of the harmonic system in total hip arthroplasty: a prospective, comparative, observational stud...