Document Detail


Long-term cardiopulmonary support with a composite artificial heart-lung system.
MedLine Citation:
PMID:  8573867     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A novel cardiopulmonary support (CPS) system as a long-term total substitute for the natural heart and lung was designed and evaluated. The CPS system is composed of a ventricular assist device (VAD) and a newly developed compact artificial lung (AL). The pneumatic VAD was originally developed in the authors' institute and has proven durable and antithrombogenic in over 120 clinical uses. The AL, with 1.2 m2 membrane surface area and 140 ml priming volume, consists of a special gas exchange membrane where not only are micropores blind-ended to prevent serum leakage, but the blood contacting surface is treated with heparin binding to ease systemic anticoagulation. Chronic animal experiments of total CPS were carried out in three goats weighing from 28 to 36 kg. Venoarterial bypass was instituted with the CPS system in which all venous blood was drawn from both the right atrium and ventricle and returned to the aorta. No systemic anticoagulation was conducted except for heparin added fluid infusion from pressure lines. One animal survived for 2 weeks without any device related problems until electively euthanized. Blood flow rate ranged between 3.2 to 4.5 L/min, O2 transfer between 120 to 220 ml/min, and CO2 removal between 60 to 210 ml/min. No substantial changes were found in platelet count or other coagulation factors. These results indicate that the system has the ability to be used for long-term CPS, such as a bridge to lung or heart-lung transplantation.
Authors:
E Tatsumi; K Eya; Y Taenaka; T Nakatani; Y Baba; T Masuzawa; Y Wakisaka; K Toda; K Miyazaki; H Takano
Related Documents :
11396167 - Increasing drug concentration in a rat lung tumor model by combining isolated lung perf...
23730647 - The antihypertensive efficacy of chlorthalidone and telmisartan in indian hypertensive ...
3994027 - Hypoxic pulmonary vasoconstriction in the human lung: effect of repeated hypoxic challe...
16398387 - The theoretical basis for using apnoeic oxygenation via the non-ventilated lung during ...
8096167 - Subtle features of the hemodynamic response to amyl nitrite inhalation: new aspects of ...
9104907 - Clinical application of pulsed doppler tissue imaging for assessing abnormal left ventr...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  ASAIO journal (American Society for Artificial Internal Organs : 1992)     Volume:  41     ISSN:  1058-2916     ISO Abbreviation:  ASAIO J.     Publication Date:    1995 Jul-Sep
Date Detail:
Created Date:  1996-03-12     Completed Date:  1996-03-12     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  9204109     Medline TA:  ASAIO J     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  M557-60     Citation Subset:  IM    
Affiliation:
Department of Artificial Organs, National Cardiovascular Center Research Institute, Osaka, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Artificial Organs*
Blood Coagulation
Blood Flow Velocity
Carbon Dioxide / blood
Evaluation Studies as Topic
Extracorporeal Membrane Oxygenation
Goats
Heart, Artificial*
Heart-Assist Devices
Heart-Lung Machine
Hemodynamics
Humans
Lung*
Oxygen / blood
Time Factors
Chemical
Reg. No./Substance:
124-38-9/Carbon Dioxide; 7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  An atraumatic aspirator for use in autologous transfusion and cardiac bypass.
Next Document:  Evaluation of a new heparin bonding process in prolonged extracorporeal membrane oxygenation.