Document Detail


Left intraventricular balloon pump optimization during intractable cardiac arrest.
MedLine Citation:
PMID:  8841857     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
This work aims to determine optimal balloon shape and volume during left intraventricular balloon pumping (IABP) in the fibrillating dog heart. A balloon volume equal to the left ventricular end-diastolic volume (LVEDV) maintained a higher systolic aortic pressure and flow (106.4 +/- 2.7 mmHg and 84.7 +/- 2.35 ml/kg/min, x +/- SEM, respectively) than a 25% smaller (97.8 +/- 3.3 mmHg, P = 0.002 and 63.7 +/- 4.1 ml/kg/min, P = 0.002, respectively) or a 25% larger balloon (87.4 +/- 2.3 mmHg, P = 0.002 and 70.9 +/- 3.4 ml/kg/min, P = 0.002, respectively). Among 5 different balloon shapes tested, a pear-shaped balloon inflated from the apex to the base of the left ventricle induced the highest (P varying from 0.042 to 0.01, compared to the remaining balloon shapes) systolic aortic pressure and flow (104.6 +/- 4.5 mmHg and 77.9 +/- 1.7 mg/kg/min, respectively). In conclusion, a pear shaped balloon, inflated to a volume equal to the LVEDV, from the apex to the base of the left ventricle, induced an optimal hemodynamic effect during LVBP.
Authors:
S F Stamatelopoulos; L Kochilas; N S Saridakis; N A Zakopoulos; S D Moulopoulos
Related Documents :
12889397 - Intermittent pneumatic compression pump settings for the optimum reduction of oedema.
1584057 - Is endothelium necessary for transmural pressure-induced contractions of bovine truncal...
2597547 - Left ventricular assist system with a simplified cannulation technique.
8728857 - The heart of the icefish: bioconstruction and adaptation.
9260837 - Pressure dependence of so-called transtrophoblastic channels during fetal perfusion of ...
25181847 - Fourier analysis of peripheral blood pressure and flow in intraoperative assessment of ...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  The International journal of artificial organs     Volume:  19     ISSN:  0391-3988     ISO Abbreviation:  Int J Artif Organs     Publication Date:  1996 Jul 
Date Detail:
Created Date:  1996-12-12     Completed Date:  1996-12-12     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  7802649     Medline TA:  Int J Artif Organs     Country:  ITALY    
Other Details:
Languages:  eng     Pagination:  422-30     Citation Subset:  IM    
Affiliation:
Department of Clinical Therapeutics, Alexandra General Hospital, Athens University Medical School, Greece.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Analysis of Variance
Animals
Blood Pressure / physiology
Disease Models, Animal
Dogs
Echocardiography
Heart Arrest / therapy*
Heart-Assist Devices / standards*
Intra-Aortic Balloon Pumping*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Extracorporeal artificial liver: the influence of a second cell layer on the morphology and function...
Next Document:  The Fresenius continuous autotransfusion system (CATS): preliminary studies and application.