K-edge digital subtraction arthrography of the painful hip prosthesis: a feasibility study.
MedLine Citation:
PMID:  2120935     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
K-edge energy subtraction radiography is a method for detecting the presence of iodinated contrast material by subtracting two digital radiographs produced by X-ray beams with energies above and below the iodine K edge. We performed a feasibility study on the application of K-edge energy digital subtraction arthrography (KEDSA) to painful hip prostheses. During arthrography, loosening of the prosthesis is implied if contrast material is seen dissecting around the prosthesis, an often difficult detection task because of adjacent prosthesis metal or cement. In conventional arthrography a preliminary mask image is thus used from which films obtained after injection of iodinated contrast material are subtracted. Movement by the patient during this process may preclude subsequent subtraction. With KEDSA, since multiple image pairs may be obtained after the injection of contrast material, the problem of patient motion is virtually eliminated. A conventional X-ray tube operating between 55 and 65 kVp was alternately filtered by iodine and cerium filters to produce the KEDSA images. The apparatus was capable of producing a subtracted image within 3 sec. The technique was applied to phantoms and to six patients immediately after hip arthrography that had been positive for prosthesis loosening. Although of lower spatial resolution, the KEDSA images were, in all cases, positive for loosening in a pattern consistent with the conventional arthrographic images. KEDSA was shown to be successful in detecting extraarticular contrast material. During a single study, subtraction in various imaging planes as well as postexercise subtraction imaging can be accomplished-techniques not heretofore possible in routine subtraction arthrography.
Authors:
F Kelcz; W W Peppler; C A Mistretta; A DeSmet; A A McBeath
Related Documents :
3931025 - Assessment of laryngeal function by digital subtraction laryngography.
18033945 - Efficacy of a new software in eliminating the angulation errors in digital subtraction ...
3313155 - Image enhancement of the mandibular condyle through digital subtraction.
22669455 - Use of panitumumab-irdye800 to image microscopic head and neck cancer in an orthotopic ...
18222815 - Automatic segmentation of ultrasound images using morphological operators.
20798785 - A data-driven approach to prior extraction for segmentation of left ventricle in cardia...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  AJR. American journal of roentgenology     Volume:  155     ISSN:  0361-803X     ISO Abbreviation:  AJR Am J Roentgenol     Publication Date:  1990 Nov 
Date Detail:
Created Date:  1990-11-19     Completed Date:  1990-11-19     Revised Date:  2008-02-15    
Medline Journal Info:
Nlm Unique ID:  7708173     Medline TA:  AJR Am J Roentgenol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1053-8     Citation Subset:  AIM; IM    
Affiliation:
Department of Radiology, University of Wisconsin Hospitals and Clinics, Madison 53792.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Aged, 80 and over
Arthrography / methods
Computer Simulation
Contrast Media
Feasibility Studies
Female
Hip Joint / radiography*
Hip Prosthesis*
Humans
Male
Middle Aged
Models, Structural
Pain / etiology
Prosthesis Failure
Subtraction Technique*
Chemical
Reg. No./Substance:
0/Contrast Media