Investigation of geometric distortions on magnetic resonance and cone beam computed tomography images used for planning and verification of high-dose rate brachytherapy cervical cancer treatment.
MedLine Citation:
PMID:  20149759     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To measure the amount of geometric distortions present in the three-dimensional imaging modalities--cone beam computed tomography (CBCT) and magnetic resonance imaging (MRI)--used at University of California, San Francisco, CA, for gynecologic high dose rate brachytherapy.
METHODS AND MATERIALS: An MRI- and CT-compatible water phantom with two different sets of support structures was designed and built for this study. The support structures were used to precisely position catheters that were filled with either an MRI contrast agent or a string of radio-opaque markers. The first support structure without anatomy was built to test system-based distortions. A second structure included two types of gynecologic applicators as well as several anatomical structures, including bones and rectum to test object-induced distortions. Images were acquired with CT (for reference), kilovoltage CBCT, and MRI (1.5 T with T1- and T2-weighted images). The difference in catheter positions between the images and the CT images was analyzed.
RESULTS: For CBCT, the mean of the absolute deviations was below 1mm in all directions. The inherent uncertainty in the measurement of distortion was less than 0.5mm. MRI presented mean absolute system-based distortions between 0.6 and 1.1mm in the central region of the image and between 0.7 and 2.3mm in the outer region. Images with the applicator and anatomy in place created mean absolute distortions of 0.4, 0.8, and 0.8mm or less for CBCT, MR-T1, and MR-T2 images, respectively.
CONCLUSIONS: The distortions measured in the presence of applicators are small enough to validate the use of CBCT and 1.5 T MRI for GYN brachytherapy treatment planning and verification.
Authors:
Jean-François Aubry; Joey Cheung; Olivier Morin; Luc Beaulieu; I-Chow Hsu; Jean Pouliot
Related Documents :
12924459 - Ct verification of isocentre relocatability using stereotactic mask fixation system.
16024179 - A novel 3d volumetric voxel registration technique for volume-view-guided image registr...
10494139 - Electrophysiological versus image-based targeting in the posteroventral pallidotomy.
22005909 - Splenic infarction following conventional open gastrectomy in patients with gastric mal...
21750139 - Development of a risk score to guide brain imaging in older patients admitted with fall...
17077649 - Direct anatomical localization of the subthalamic nucleus on ct with comparison to scha...
19422919 - New mri, 18f-dopa and 11c-(+)-alpha-dihydrotetrabenazine templates for macaca fascicula...
21042209 - Dissociation between neurovegetative signs and subjective symptoms in a case of idiopat...
12202709 - Detection of hepatic metastases from cancers of the gastrointestinal tract by using non...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't     Date:  2010-02-09
Journal Detail:
Title:  Brachytherapy     Volume:  9     ISSN:  1873-1449     ISO Abbreviation:  Brachytherapy     Publication Date:    2010 Jul-Sep
Date Detail:
Created Date:  2010-07-26     Completed Date:  2010-12-15     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101137600     Medline TA:  Brachytherapy     Country:  United States    
Other Details:
Languages:  eng     Pagination:  266-73     Citation Subset:  IM    
Copyright Information:
(c) 2010 American Brachytherapy Society. Published by Elsevier Inc. All rights reserved.
Affiliation:
Department of Radiation Oncology, University of California San Francisco, San Francisco, CA 94143, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Artifacts*
Cone-Beam Computed Tomography / instrumentation,  methods*
Female
Humans
Image Enhancement / methods
Image Interpretation, Computer-Assisted / methods*
Magnetic Resonance Imaging / instrumentation,  methods*
Phantoms, Imaging
Radiotherapy Planning, Computer-Assisted / methods*
Reproducibility of Results
Sensitivity and Specificity
Uterine Cervical Neoplasms / diagnosis*,  radiotherapy*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Prostate gland edema after single-fraction high-dose rate brachytherapy before external beam radiati...
Next Document:  Identification and quantitative determination of a major circulating metabolite of gambogic acid in ...