Document Detail


Insulation Failure with Externalized Conductor of a Linox SD Lead: A Case Report.
MedLine Citation:
PMID:  24383881     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
A 53-year-old man with a dual-chamber implantable cardioverter-defibrillator (ICD) presented to the electrophysiology (EP) lab for an EP study in evaluation of supra-ventricular tachycardia despite medical therapy. Fluoroscopic imaging during placement of diagnostic catheters demonstrated insulation failure with an externalized conductor on the right ventricular defibrillator lead just distal to the proximal defibrillator coil. Device interrogation revealed normal pacing, and shock impedances, normal sensing, and stable pacing thresholds. This article is protected by copyright. All rights reserved.
Authors:
Joseph A Manfredi; Steven M Smithgall; Charles M Kircher; Mary Alice Lollis
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-1-2
Journal Detail:
Title:  Journal of cardiovascular electrophysiology     Volume:  -     ISSN:  1540-8167     ISO Abbreviation:  J. Cardiovasc. Electrophysiol.     Publication Date:  2014 Jan 
Date Detail:
Created Date:  2014-1-3     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9010756     Medline TA:  J Cardiovasc Electrophysiol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
This article is protected by copyright. All rights reserved.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  New mini- zincin structures provide a minimal scaffold for members of this metallopeptidase superfam...
Next Document:  The renaissance of bacillosamine and its derivatives: pathway characterization and implications in p...