Document Detail


Instantaneous exposure to nuclear medicine staff involved in PET-CT imaging in developing countries: experience from a tertiary care centre in India.
MedLine Citation:
PMID:  22246804     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
PURPOSE: To assess the radiation exposure to nuclear medicine staff at a positron emission tomography-computed tomography (PET-CT) centre with high patient throughput. MATERIALS AND METHODS: This prospective study included 70 adult patients who underwent 18F-fluorodeoxyglucose (FDG) PET-CT for their clinical indications. The patients' actual injected FDG activity was calculated by subtracting the syringe activity (post-injection) from the loaded syringe activity (pre-injection). The instantaneous exposure to nuclear medicine staff involved in PET-CT imaging was measured. The instantaneous dose rate of the physicians was recorded during FDG injection and that of the technologist was recorded during the patient's positioning, respectively, at 1.0-m distance from the anterior chest using a calibrated portable gamma-ray survey meter. RESULTS: The average FDG activity injected in adult patients was 308.5 MBq (range 173.1-438.8 MBq). The instantaneous exposure to the nuclear medicine (NM) physician during the injection time was 31 μSv/h (14-60 μSv/h). The instantaneous exposure to the NM technologist during positioning was 18 (10-34) μSv/h. With an average of 10 patients per day, the quarterly dose to physicians was 628 μSv and to technologists 182 μSv for 300 patients. The extrapolated annual dose was 2.5 mSv for physicians and 0.7 mSv for technologists, respectively. CONCLUSION: Instantaneous exposure of nuclear medicine staff involved in PET-CT imaging at a busy tertiary care centre is within permissible limits of the International Commission on Radiological Protection (ICRP-103) (total <100 mSv over 5 years and not >50 mSv in a single year) and atomic energy regulatory board (total <100 mSv over 5 years and not >30 mSv in a single year).
Authors:
Sunil Kumar; Anil Kumar Pandey; Punit Sharma; Shamim Ahmed Shamim; Arun Malhotra; Rakesh Kumar
Related Documents :
15821964 - Hyperpolarized 3he mri and 81mkr spect in chronic obstructive pulmonary disease.
19542254 - Qualitative analysis of high-resolution ct scans in severe asthma.
22016114 - Computed tomography fluoroscopy-guided radiofrequency ablation following intra-arterial...
11693664 - Irregularity of airway branching in a mouse bronchial tree: a 3-d morphometric study.
14696224 - Echo planar mr imaging of the liver: comparison of images with and without motion probi...
16274894 - Chronic, reactive conditions of the oral cavity simulating mucosal carcinomas: ct and m...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-1-14
Journal Detail:
Title:  Japanese journal of radiology     Volume:  -     ISSN:  1867-108X     ISO Abbreviation:  -     Publication Date:  2012 Jan 
Date Detail:
Created Date:  2012-1-16     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101490689     Medline TA:  Jpn J Radiol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Department of Nuclear Medicine, All India Institute of Medical Sciences, E-81, Ansari Nagar (East), AIIMS Campus, New Delhi, 110029, India.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Flexible transceiver array for ultrahigh field human MR imaging.
Next Document:  Saturated fatty acid induction of endoplasmic reticulum stress and apoptosis in human liver cells vi...