Document Detail


Inspiratory valve malfunction in a circle system: pitfalls in capnography.
MedLine Citation:
PMID:  1451230     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Capnography is a useful technique in monitoring the integrity of anaesthetic equipment such as the malfunctioning of unidirectional valves in circle system. However, the lack of a precise mechanism in existing capnographs to identify the start of inspiration and the beginning of expiration in the capnograms, makes the analysis of the carbon dioxide waveforms during inspiration difficult and thus results in inaccurate assessment of rebreathing. We report a case where, during the malfunction of the inspiratory unidirectional valve in the circle system, the capnograph failed to detect the presence of substantial rebreathing. Critical analysis of the capnogram recorded during the malfunction revealed that there was substantial rebreathing which was underestimated by the capnograph as it reports only the lowest CO2 concentration rebreathed during inspiration in such abnormal situations.
Authors:
A Y Kumar; K Bhavani-Shankar; H S Moseley; Y Delph
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Canadian journal of anaesthesia = Journal canadien d'anesthésie     Volume:  39     ISSN:  0832-610X     ISO Abbreviation:  Can J Anaesth     Publication Date:  1992 Nov 
Date Detail:
Created Date:  1992-12-31     Completed Date:  1992-12-31     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  8701709     Medline TA:  Can J Anaesth     Country:  CANADA    
Other Details:
Languages:  eng     Pagination:  997-9     Citation Subset:  IM    
Affiliation:
Department of Anaesthesia, Queen Elizabeth Hospital, University of West Indies, Faculty of Medical Sciences, Barbados.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Anesthesia, Closed-Circuit / adverse effects*,  instrumentation*
Carbon Dioxide / analysis*
Equipment Failure
Female
Humans
Inhalation
Middle Aged
Monitoring, Physiologic / instrumentation,  methods*
Pressure
Pulmonary Ventilation / physiology
Respiration, Artificial
Tidal Volume
Chemical
Reg. No./Substance:
124-38-9/Carbon Dioxide

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Arterial baroreflex attenuation during and after continuous propofol infusion.