| Innovations in shock wave lithotripsy technology: updates in experimental studies. | |
| | |
MedLine Citation:
|
PMID: 15540748 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
PURPOSE: We developed innovations in shock wave lithotripsy (SWL) technology. MATERIALS AND METHODS: Two technical upgrades were implemented in an original unmodified HM-3 lithotriptor (Dornier Medical Systems, Inc., Kennesaw, Georgia). First, a single unit ellipsoidal reflector insert was used to modify the profile of lithotriptor shock wave (LSW) to decrease the propensity of tissue injury in SWL. Second, a piezoelectric annular array (PEAA) generator (f = 230 kHz and F = 150 mm) was used to produce an auxiliary shock wave of approximately 13 MPa in peak pressure (at 4 kV output voltage) to intensify the collapse of LSW induced bubbles near the target stone for improved comminution efficiency. RESULTS: Consistent rupture of a vessel phantom made of single cellulose hollow fiber (i.d. = 0.2 mm) was produced after 30 shocks by the original HM-3 reflector at 20 kV. In comparison no vessel rupture could be produced after 200 shocks using the upgraded reflector at 22 kV or the PEAA generator at 4 kV. Using cylindrical BegoStone phantoms (Bego USA, Smithfield, Rhode Island) stone comminution efficiencies (mean +/- sd) after 1,500 shocks produced by the original and upgraded HM-3 reflectors, and the combined PEAA/upgraded HM-3 system, were 81.3% +/- 3.5%, 90.1% +/- 4.3% and 95.2% +/- 3.3%, respectively (p<0.05). CONCLUSIONS: Optimization of the pulse profile and sequence of LSW can significantly improve stone comminution while simultaneously decreasing the propensity of tissue injury during in vitro SWL. This novel concept and associated technologies may be used to upgrade other existing lithotriptors and to design new shock wave lithotriptors for improved performance and safety. |
| | |
Authors:
|
Yufeng Zhou; Franklin H Cocks; Glenn M Preminger; Pei Zhong |
Related Documents
:
|
11311668 - Telemedicine in china. 15288208 - Artificial parthenogenesis and control of voltinism to manage transgenic populations in... 18649968 - Considering something 'else': ethical, legal and socio-economic factors in medical imag... 8270838 - Object-oriented data management: an approach to computerized anaesthesia documentation. 10525538 - Dynactin is required for microtubule anchoring at centrosomes. 12527158 - Does smoking affect the outcome of patient education and self-management in asthmatics? |
Publication Detail:
|
Type: Journal Article; Research Support, U.S. Gov't, P.H.S. |
Journal Detail:
|
Title: The Journal of urology Volume: 172 ISSN: 0022-5347 ISO Abbreviation: J. Urol. Publication Date: 2004 Nov |
Date Detail:
|
Created Date: 2004-11-15 Completed Date: 2005-01-06 Revised Date: 2007-11-14 |
Medline Journal Info:
|
Nlm Unique ID: 0376374 Medline TA: J Urol Country: United States |
Other Details:
|
Languages: eng Pagination: 1892-8 Citation Subset: AIM; IM |
Affiliation:
|
Department of Mechanical Engineering and Materials Science, Duke University, Durham, North Carolina 27708-0300, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Equipment Design Lithotripsy / instrumentation* |
| Grant Support | |
ID/Acronym/Agency:
|
R01-DK52985/DK/NIDDK NIH HHS; R01-DK58266/DK/NIDDK NIH HHS |
| Comments/Corrections | |
Comment In:
|
J Urol. 2004 Nov;172(5 Pt 1):1774
[PMID:
15540719
]
|
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Comparison of 2 generations of piezoelectric lithotriptors using matched pair analysis.