Document Detail


Innovations in intra-aortic balloon pump management: computer modem technology.
MedLine Citation:
PMID:  10150680     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The computer modem technology being developed by Datascope Corporation for the purpose of off-site intra-aortic balloon pump (IABP) monitoring was clinically evaluated in 30 patients requiring IABP support. The Datascope System 95 model IABP with built-in modem was used on all patients. Remote communications via a personal computer with modem were conducted under both routine and emergency settings. The Datascope PC IABP software was evaluated for its user compatibility, efficiency, diagnostic capability and overall usefulness as a clinical tool. During the eight month evaluation period, 87 remote communications were conducted for both routine and emergency IABP evaluation checks. Adjustments were recommended on 22 occasions relevant to balloon timing, trigger mode selection and augmentation volume settings. Eight communications were initiated in emergency settings due to a variety of patient conditions. Emergency intervention was successful in diagnosing and resolving critical situations including atrial arrhythmias, pacemaker timing, low cardiac output syndrome, loss of trigger source, catheter malpositioning and poor augmentation. The diagnostic capabilities and efficient means of data collection by the computer software provide the clinician with a valuable tool for routine IABP clinical monitoring, as well as emergency problem resolution.
Authors:
J M Ingram
Publication Detail:
Type:  Clinical Trial; Journal Article    
Journal Detail:
Title:  The Journal of extra-corporeal technology     Volume:  26     ISSN:  0022-1058     ISO Abbreviation:  J Extra Corpor Technol     Publication Date:  1994 Sep 
Date Detail:
Created Date:  1995-09-21     Completed Date:  1995-09-21     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  0267637     Medline TA:  J Extra Corpor Technol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  135-9     Citation Subset:  T    
Affiliation:
Cedars Medical Center, Miami, Florida, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Humans
Intra-Aortic Balloon Pumping*
Local Area Networks
Microcomputers
Middle Aged
Monitoring, Physiologic / methods
Signal Processing, Computer-Assisted*
Software Validation

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Nafamostat mesilate reduces blood cell adhesion to cardiopulmonary bypass circuits: an in-vitro stud...