Document Detail


Initial experience with the Cardiva Boomerang vascular closure device in diagnostic catheterization.
MedLine Citation:
PMID:  17195965     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: The authors studied the safety and efficacy of the Cardiva Boomerang vascular closure device in patients undergoing diagnostic cardiac catheterization. BACKGROUND: Conventional vascular closure devices (sutures, collagen plugs, or metal clips) have been associated with catastrophic complications including arterial occlusion and foreign body infections; furthermore, they cannot be utilized in patients with peripheral vascular disease or vascular access site in a vessel other than the common femoral artery. The Cardiva Boomerang device facilitates vascular hemostasis without leaving any foreign body behind at the access site, can be used in peripheral vascular disease, and can be used in vessels other than the common femoral artery METHODS: A total of 96 patients undergoing transfemoral diagnostic cardiac catheterization were included in this study, including 25 (26%) patients with contraindications to conventional closure devices. Femoral angiography was performed prior to deployment of the Cardiva Boomerang closure device. Patients were ambulated at 1 hr after hemostasis was achieved. RESULTS: The device was successfully deployed and hemostasis achieved with the device alone in 95 (99%) patients. The device failed to deploy in 1 (1%) patient and required conversion to standard manual compression. Minor complications were observed in 5 (5%) patients. No patients experienced major complications including femoral hematoma > 4 cm, red blood cell transfusion, retroperitoneal bleed, arteriovenous fistula, pseudoaneurysm, infection, arterial occlusion, or vascular surgery. CONCLUSIONS: The Cardiva Boomerang device is safe and effective in patients undergoing diagnostic cardiac catheterization using the transfemoral approach, facilitating early ambulation with low rates of vascular complications.
Authors:
Brendan J Doyle; Michael J Godfrey; Ryan J Lennon; James L Ryan; John F Bresnahan; Charanjit S Rihal; Henry H Ting
Related Documents :
1533945 - Peripheral directional atherectomy: 4-year experience.
8421335 - Risks and benefits of femoropopliteal percutaneous balloon angioplasty.
23422155 - A preliminary study of the effects of ulinastatin on early postoperative cognition func...
11434895 - Intermittent claudication: effective medical management of a common circulatory problem.
9795915 - Prolonged treatment with interferon in patients with histologically mild chronic hepati...
23464955 - Clavicular length: the assumption of symmetry.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions     Volume:  69     ISSN:  1522-1946     ISO Abbreviation:  Catheter Cardiovasc Interv     Publication Date:  2007 Feb 
Date Detail:
Created Date:  2007-01-31     Completed Date:  2007-04-03     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100884139     Medline TA:  Catheter Cardiovasc Interv     Country:  United States    
Other Details:
Languages:  eng     Pagination:  203-8     Citation Subset:  IM    
Copyright Information:
(c) 2006 Wiley-Liss, Inc.
Affiliation:
Division of Cardiovascular Diseases, Mayo Clinic, Rochester, Minnesota 55905, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Equipment Design
Equipment Safety
Female
Femoral Artery / surgery*
Heart Catheterization*
Heart Diseases / diagnosis*
Hemostatic Techniques / instrumentation*
Humans
Male
Middle Aged
Punctures / instrumentation
Treatment Outcome

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The impact of renal insufficiency on patients' outcomes in emergent angioplasty for acute myocardial...