Initial evaluation of robotic technology for microsurgical vasovasostomy.
MedLine Citation:
PMID:  14665899     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: Conventional microscope assisted vasovasostomy (MAVV) is a technically difficult procedure that is most successful in the hands of well-trained microsurgeons. Robotics may help surgeons overcome the microsurgical challenges of tremor, limited dexterity, miniaturized instrumentation and use of fine suture. We determine the feasibility of a robotic assisted vasovasostomy (RAVV) and compare performance measures with those of conventional MAVV. MATERIALS AND METHODS: One surgeon performed 10 vasovasostomies with a modified 1-layer technique and 9-zero suture on fresh human vas specimens using the robot in 5 RAVV cases and standard microsurgical instrumentation in 5 MAVV cases. Pre-specified performance measures and adverse haptic events (broken sutures, bent needles or loose stitches) were recorded. Patency was evaluated by instilling saline through the anastomoses. RESULTS: Mean operating time and number of adverse haptic events were higher for RAVV than for MAVV (84 vs 38 minutes, p = 0.01; 2.4 vs 0.0 events, p = 0.03). The number of needle passes required for the 6 full-thickness stitches was similar in both groups (16.8 vs 15.2 passes, p = 0.55). Although no tremor occurred during RAVV, minimal to moderate amounts occurred during MAVV. Minimal fatigue was noted for both groups. Patency was confirmed in all 10 operations. CONCLUSIONS: Use of RAVV in this human ex vivo vas model was feasible. While RAVV took longer to perform and was associated with adverse haptic events, elimination of tremor and comparable patency rates suggest that it may be a viable surgical alternative for microsurgical vasovasostomy.
Authors:
Wayne Kuang; Paul R Shin; Surena Matin; Anthony J Thomas
Related Documents :
2222389 - The agoraphobia scale: an evaluation of its reliability and validity.
11220369 - Quantitative and qualitative outcome measures after thalamic deep brain stimulation to ...
24968509 - The effects of a unilateral gluteal activation protocol on single leg drop jump perform...
9155809 - Early dentofacial features of class ii malocclusion: a longitudinal study from the deci...
23049549 - The feasibility and validity of a remote pulse oximetry system for pulmonary rehabilita...
18602219 - Modifications of the obsessive compulsive drinking scale (ocds-g) for use in longitudin...
20142779 - Harmful effect of land-based endurance exercise in rats with diabetic nerve.
8729469 - Pulmonary oxygen uptake kinetics in nonsteady state.
24947029 - Cortisol is not the primary mediator for augmented cxcr4 expression on natural killer c...
Publication Detail:
Type:  Comparative Study; In Vitro; Journal Article    
Journal Detail:
Title:  The Journal of urology     Volume:  171     ISSN:  0022-5347     ISO Abbreviation:  J. Urol.     Publication Date:  2004 Jan 
Date Detail:
Created Date:  2003-12-10     Completed Date:  2004-01-29     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0376374     Medline TA:  J Urol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  300-3     Citation Subset:  AIM; IM    
Affiliation:
Glickman Urological Institute, The Cleveland Clinic Foundation, Ohio 44195, USA. kuangw@ccf.org
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Equipment Design
Feasibility Studies
Humans
Male
Microsurgery
Robotics* / instrumentation
Suture Techniques
Vasovasostomy / methods*
Comments/Corrections
Comment In:
J Urol. 2004 Aug;172(2):780; author reply 780-1   [PMID:  15247793 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Questionable efficacy of extracorporeal shock wave therapy for Peyronie's disease: results of a pros...
Next Document:  The impact of obstructive interval and sperm granuloma on outcome of vasectomy reversal.