Document Detail


Infusion pump delivers over-dosage of propofol as a result of missing syringe support.
MedLine Citation:
PMID:  16551915     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
We describe the malfunction of a common drug infusion pump. The syringe saddle was missing and allowed the syringe barrel to contact the pump case, which decreased the outward displacement of the syringe clamp. Then, the infusion pump falsely detected a smaller syringe size and consequently delivered an increased infusion rate and overdose of propofol to the patient. More commonly, an incorrectly mounted syringe may increase the outward displacement of the syringe clamp so that the infusion pump falsely detects a larger syringe size, with resultant less than expected infusion rate.
Authors:
Christian Koch; Christine Hollister; Peter H Breen
Related Documents :
8712445 - Acute pain and central nervous system arousal do not restore impaired hypoxic ventilato...
9587285 - Evaluation of lidocaine in human inferior alveolar nerve block.
3337795 - Comparison of the effects of subanaesthetic concentrations of isoflurane or nitrous oxi...
2572135 - Reversal of profound paralysis: use of large doses of edrophonium to antagonize vecuron...
6734215 - Telemetric evaluation of the atrioventricular conduction system.
6241785 - Ventricular fibrillation threshold in the three-toed sloth (bradypus tridactylus).
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  102     ISSN:  1526-7598     ISO Abbreviation:  Anesth. Analg.     Publication Date:  2006 Apr 
Date Detail:
Created Date:  2006-03-22     Completed Date:  2006-04-13     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1154-6     Citation Subset:  AIM; IM    
Affiliation:
Department of Anesthesiology, University of California-Irvine, Orange, California, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Drug Delivery Systems / adverse effects*,  instrumentation*
Equipment Failure
Female
Humans
Infusion Pumps / adverse effects*
Infusions, Intravenous
Middle Aged
Overdose
Propofol / adverse effects*
Syringes / adverse effects*
Chemical
Reg. No./Substance:
2078-54-8/Propofol

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The delivery of drugs to patients by continuous intravenous infusion: modeling predicts potential do...