Document Detail


Influence of xylazine on the function of the LiDCO sensor in isoflurane anaesthetized horses.
MedLine Citation:
PMID:  24890704     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: Previous studies showed an influence of xylazine on the LiDCO sensor in vitro and in standing horses, but did not prove that this interaction caused error in LiDCO measurements. Therefore, agreement of cardiac output (CO) measurements by LiDCO and bolus-thermodilution (BTD) was determined in horses receiving xylazine infusions.
STUDY DESIGN: Prospective, experimental study.
ANIMALS: Eight Warmblood horses.
METHODS: All horses were premedicated with xylazine. Anaesthesia was induced with midazolam and ketamine and was maintained with isoflurane in oxygen. During six hours of anaesthesia CO measurements and blood samples were taken before, during and after a 60 minute period of xylazine infusion. Pairs of LiDCO and bolus thermo-dilution (BTD) measurements of CO were performed. Sensor voltages exposed to blood and saline were measured before, during and after xylazine infusion and compared using Bland-Altman method of agreement with corrections for repeated measures.
RESULTS: The CO values (mean ± SD) before xylazine were 34.8 ± 7.3 and 36.4 ± 8.1 L minute(-1) for BTD and LiDCO, respectively. After starting the xylazine infusion, the CO values for BTD decreased to 27.5 ± 6.1 L minute(-1) whereas CO values measured by LiDCO increased to 54.7 ± 18.4 L minute(-1) . One hour after discontinuing xylazine infusion, CO values were 33 ± 6.7 and 36.5 ±11.9 L minute(-1) for BTD and LiDCO, respectively. The difference between saline and blood exposed sensor voltages decreased during xylazine infusion and these differences were positive numbers before but negative during the infusion. There were correlations between xylazine plasma concentrations, CO differences and sensor voltage differences (saline - blood).
CONCLUSIONS AND CLINICAL RELEVANCE: This study proved that xylazine infusion caused concentration dependent bias in LiDCO measurements leading to an overestimation of readings. Sensor voltage differences (saline - blood) may become valuable clinical tool to predict drug-sensor interactions.
Authors:
Klaus Hopster; Tamas D Ambrisko; Jessica Stahl; Johannes P Schramel; Sabine Br Kästner
Related Documents :
17132644 - Predictive performance of 'servin's formula' during bis-guided propofol-remifentanil ta...
7917504 - Inhaled tert-butyl acetate and its metabolite tert-butyl alcohol accumulate in the bloo...
17868264 - The relationship of methanol and formate concentrations in fatalities where methanol is...
8452194 - Anticipated stimulant and sedative effects of alcohol vary with dosage and limb of the ...
2029254 - An improved method for the purification of blood cells of turkeys.
20583274 - Data analysis in flow cytometry: the future just started.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-5-29
Journal Detail:
Title:  Veterinary anaesthesia and analgesia     Volume:  -     ISSN:  1467-2995     ISO Abbreviation:  Vet Anaesth Analg     Publication Date:  2014 May 
Date Detail:
Created Date:  2014-6-3     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100956422     Medline TA:  Vet Anaesth Analg     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
© 2014 Association of Veterinary Anaesthetists and the American College of Veterinary Anesthesia and Analgesia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Versatile types of polysaccharide-based supramolecular polycation/pDNA nanoplexes for gene delivery.
Next Document:  Design, synthesis, and evaluation of 7H-thiazolo[3,2-b]-1,2,4-triazin-7-one derivatives as dual bind...