Influence of various instrument settings on the flow information derived from the power mode.
MedLine Citation:
PMID:  2021011     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
To evaluate the factors influencing flow information displayed by the power operating mode, color Doppler flow mapping was used to image the flow in an in vitro left heart pulse duplicator system. The effect of changing one instrument setting, such as pulse repetition frequency, color filter and frame rate while keeping all other instrumental settings constant, was noted on the displayed flow areas. The flow areas derived using power and velocity modes were also compared. An increase in pulse repetition frequency and color filter decreased the flow areas significantly, and a flow area increase occurred with a decrease in frame rate. No significant difference was observed between the flow areas derived using the two operating modes. Like the velocity mode, the power mode display is also influenced by instrument settings. Although low velocity flows are better delineated using this mode, however, no significant difference occurred in the flow areas measured by this mode and velocity mode. Further studies need to be conducted to address its potential applications in the clinical setting and in quantitation.
Authors:
S P Jain; P H Fan; E F Philpot; N C Nanda; K K Aggarwal; S Moos; A P Yoganathan
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Ultrasound in medicine & biology     Volume:  17     ISSN:  0301-5629     ISO Abbreviation:  Ultrasound Med Biol     Publication Date:  1991  
Date Detail:
Created Date:  1991-05-29     Completed Date:  1991-05-29     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  0410553     Medline TA:  Ultrasound Med Biol     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  49-54     Citation Subset:  IM    
Affiliation:
Heart Station, University of Alabama, Birmingham 35294.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Flow Velocity
Coronary Circulation*
Echocardiography, Doppler / methods*
Heart / physiology*
Humans
Models, Structural

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The relative contributions of myocardial wall thickness and ischemia to ultrasonic myocardial integr...
Next Document:  Ultrasound effects on cell proliferation and migration of chick motoneurons.