Influence of PET reconstruction parameters on the TrueX algorithm. A combined phantom and patient study.
MedLine Citation:
PMID:  23348719     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
With the increasing use of functional imaging in modern radiotherapy (RT) and the envisaged automated integration of PET into target definition, the need for reliable quantification of PET is growing. Reconstruction algorithms in new PET scanners employ point-spread-function (PSF) based resolution recovery, however, their impact on PET quantification still requires thorough investigation. Patients, material, methods: Measurements were performed on a Siemens PET/CT using an IEC phantom filled with varying activity. Data were reconstructed using the OSEM (Gauss filter) and the PSF TrueX (Gauss and Allpass filter) algorithm with all available products of iterations (i) and subsets (ss). The recovery coeffcient (RC) and threshold defining the real sphere volume were determined for all settings and compared to the clinical standard (4i21ss). PET acquisitions of eight lung patients were reconstructed using all algorithms with 4i21ss . Volume size and tracer uptake were determined with different segmentation methods. Results: The threshold for the TrueX was lower (up to 40%) than for the OSEM. The RC for the different algorithms and filters varied. TrueX was more sensitive to permutations of i and ss and only the RC of the OSEM stabilised with increasing number. For patient scans the difference of the volume and activity between TrueX and OSEM could be reduced by applying an adapted threshold and activity correction. Conclusion: The TrueX algorithm results in excellent diagnostic image quality, however, guidelines for native algorithms have to be extended for PSF based reconstruction methods. For appropriate tumour delineation, for the TrueX a lower threshold than the 42% recommended for the OSEM is necessary. These filter dependent thresholds have to be verified for different scanners prior to using them in multicenter trials.
Authors:
B Knäusl; I F Rausch; H Bergmann; R Dudczak; A Hirtl; D Georg
Related Documents :
7408169 - Computer-assisted analysis of two-dimensional electrophoreses of human lymphoid cells.
23805869 - The feasibility of nurse practitioner-performed, telementored lung telesonography with ...
11718559 - A 11.5 a single particle reconstruction of groel using eman.
16343539 - Single particle reconstructions of the transferrin-transferrin receptor complex obtaine...
17812889 - Voyager 2 in the uranian system: imaging science results.
24978229 - Optofluidic needle probe integrating targeted delivery of fluid with optical coherence ...
7736419 - Segmentation of multidimensional cardiac images.
15001979 - Implantation of the freehand system during initial rehabilitation using minimally invas...
20012209 - Handheld histology-equivalent sectioning laser-scanning confocal optical microscope for...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-1-25
Journal Detail:
Title:  Nuklearmedizin. Nuclear medicine     Volume:  52     ISSN:  0029-5566     ISO Abbreviation:  Nuklearmedizin     Publication Date:  2013 Jan 
Date Detail:
Created Date:  2013-1-25     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7609387     Medline TA:  Nuklearmedizin     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Dr. Barbara Knäusl, Medical University of Vienna - Department of Radiooncology, Währinger Gürtel 18-20, 1090 Vienna, Austria, Tel. +43/1/404 00 72 71, E-mail: barbara.knaeusl@meduniwien.ac.at.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  From "magic bullets" to specific cancer immunotherapy.
Next Document:  Deciduous vegetation (Betula glandulosa) as a biomonitor of airborne PCB contamination from a local ...