| Inappropriate implantable cardioverter-defibrillator shock due to external alternating current leak: report of two cases. | |
| | |
MedLine Citation:
|
PMID: 15763537 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Two cases are reported of inappropriate implantable cardioverter-defibrillator shocks due to external alternating current leak. Electromagnetic interference (EMI) can mimic cardiac signals and cause inappropriate implantable cardioverter-defibrillator (ICD) shocks. EMI can arise from the normal functioning of electrical appliances or from alternating current leak. The two cases had inappropriate ICD shocks due to alternating current leak from a power drill in one and a washing machine in the other. The need for detailed advice on handling electrical equipment is emphasized. |
| | |
Authors:
|
Ngai Yin Chan; Lillian Wai-Ling Ho |
Related Documents
:
|
23374667 - Malt-type parotid lymphoma - a case report and the review of the literature. 23509837 - Osteoblastoma of the maxilla and mandible: a report of 2 cases and literature review. 21310347 - Pigmented villonodular synovitis of the temporomandibular joint: case report and review... 23259107 - Sciatica from a foraminal lumbar root schwannoma: case report and review of literature. 22131437 - Current endovascular treatment options of dural venous sinus thrombosis: a review of th... 3975807 - Spinal cord and root injuries due to glass fragments and acupuncture needles. |
Publication Detail:
|
Type: Case Reports; Journal Article |
Journal Detail:
|
Title: Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology Volume: 7 ISSN: 1099-5129 ISO Abbreviation: Europace Publication Date: 2005 Mar |
Date Detail:
|
Created Date: 2005-03-14 Completed Date: 2005-07-21 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 100883649 Medline TA: Europace Country: England |
Other Details:
|
Languages: eng Pagination: 193-6 Citation Subset: IM |
Affiliation:
|
Princess Margaret Hospital, Department of Medicine and Geriatrics, 2-10 Princess Margaret Hospital Road, Lai Chi Kok Kowloon, Hong Kong. nywlchan@yahoo.com |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Adult Defibrillators, Implantable / adverse effects* Electricity / adverse effects* Electrocardiography Equipment Failure Humans Male Middle Aged Tachycardia, Ventricular / therapy* |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Electrical storm is an independent predictor of adverse long-term outcome in the era of implantable ...