Document Detail


Inaccuracy of home sphygmomanometers: a perspective from clinical practice.
MedLine Citation:
PMID:  21928543     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
OBJECTIVE: Home blood pressure monitoring requires a precise blood pressure measurement technique and an accurate sphygmomanometer. The oscillometric sphygmomanometers have some drawbacks including in accuracy. The aims of this study were to assess the inaccuracy of home sphygmomanometers using a method first proposed in 2009 and to compare validated devices with other devices.
MATERIALS AND METHODS: Three hundred and eighty-two home sphygmomanometers that had been brought by individuals to the University Hospital were included in this study. This method has three stages(general evaluation, quick screening, final assessment) and sequential measurement of blood pressure is used for assessment. The devices were categorized into two groups for comparisons: validated and 'others'.
RESULTS: One hundred and fourteen (30%) of the 382 devices were validated. Final assessment showed 91 (24%) of the 382 devices to be inaccurate. Inaccuracy among the 'others' and wrist devices were more common than for the validated and upper arm devices (P < 0.0001 for both comparisons). The categorization of upper arm
Authors:
Tekin Akpolat; Türkan Aydogdu; Emre Erdem; Ahmet Karatas
Related Documents :
12233803 - The effects of harp music in vascular and thoracic surgical patients.
21962683 - Characterization of bsa unfolding and aggregation using a single-capillary viscometer a...
1952303 - Efficacy of tracheal insufflation of oxygen during oleic acid-induced pulmonary edema.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  16     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2011 Aug 
Date Detail:
Created Date:  2011-09-19     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  168-71     Citation Subset:  IM    
Affiliation:
Department of Nephrology, Ondokuz Mayis University School of Medicine, Samsun, Turkey. tekinakpolat@yahoo.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Power boost for labor. Landscape shift likely to trigger new wave of organizing in healthcare.
Next Document:  Relationship between the change in left ventricular hypertrophy and asleep blood pressure after sodi...