Document Detail


In vivo accuracy of two intraparenchymal intracranial pressure monitors.
MedLine Citation:
PMID:  21359608     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
PURPOSE: To evaluate the in vivo accuracy of the new Pressio(®) device for intraparenchymal monitoring of intracranial pressure (ICP) versus the Codman(®) device and intraventricular measurement external ventricular drainage (EVD). METHODS: Data were collected retrospectively for 30 consecutive patients admitted into a 25-bed neurosurgical intensive care unit of a university hospital between January and December 2009. Patients received both intraventricular and intraparenchymal ICP monitoring with Pressio(®) (n = 15) or Codman(®) (n = 15). RESULTS: We obtained 3,089 data points from the 30 patients. Mean difference between intraparenchymal and EVD pressure (bias) was -0.6 mmHg, and limits of agreement (1.96 SD of the bias) were -8.1 to 6.9 mmHg with Pressio(®) and 0.3 mmHg with limits of agreement of -6.7 to 7.1 mmHg with Codman(®) (NS). The temporal difference was -0.7 ± 1.6 mmHg/100 h of monitoring with Pressio(®) and 0.1 ± 1.6 mmHg/100 h of monitoring with Codman(®) over the study period (NS). CONCLUSIONS: Intraparenchymal pressure measured with both transducers approximates intraventricular cerebrospinal fluid pressure with an accuracy of ±7 mmHg.
Authors:
Thomas Lescot; Vincent Reina; Yannick Le Manach; Filippo Boroli; Dorian Chauvet; Anne-Laure Boch; Louis Puybasset
Related Documents :
1453728 - Does flow character of cardiopulmonary bypass make a difference?
3332268 - Duplex ultrasound monitoring of arterial grafts: prospective evaluation in conjunction ...
1998278 - Dependence of doppler echocardiographic transmitral early peak velocity on left ventric...
9080168 - High cardioplegic perfusion pressure entails reduced myocardial recovery.
21154788 - A new non-invasive method for measuring uterine vascular resistance and its relationshi...
7641338 - Automated echocardiographic measures of right ventricular area as an index of volume an...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-2-26
Journal Detail:
Title:  Intensive care medicine     Volume:  -     ISSN:  1432-1238     ISO Abbreviation:  -     Publication Date:  2011 Feb 
Date Detail:
Created Date:  2011-3-1     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7704851     Medline TA:  Intensive Care Med     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Anaesthesiology and Critical Care Department, Groupe Hospitalier Pitié-Salpêtrière, Assistance Publique-Hôpitaux de Paris et Université Pierre et Marie Curie, Paris, France, thomas.lescot@psl.aphp.fr.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Emergence of autoantibodies to HMGB1 is associated with survival in patients with septic shock.
Next Document:  Validation of Bohr dead space measured by volumetric capnography.