| In vitro performance assessment of distal protection filters: pulsatile flow conditions. | |
| | |
MedLine Citation:
|
PMID: 19995122 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
PURPOSE: To evaluate in vitro the capture efficiency of a distal protection filter (DPF) used during carotid artery stenting and examine the longitudinal vascular impedance in the presence of a DPF. METHODS: Four approved DPFs (Spider RX, FilterWire EZ, RX Accunet, and FiberNet) were evaluated in a physiologically realistic in vitro setup. A pulsatile programmable piston pump circulated a blood analog at a time-varying flow rate representative of the human common carotid artery. A silicone carotid bifurcation having average human dimensions was used for the carotid flow model. Microspheres ranging from 40 to 900 microm were injected to simulate embolization. The longitudinal vascular impedance was calculated as the ratio of the time-varying pressure gradient across the DPF to the time-varying flow rate in the internal carotid artery. RESULTS: RX Accunet had the highest capture efficiency (99.4%) and Spider RX the lowest (78.1%). Spider RX increased the longitudinal vascular impedance the least after deployment (+23%), while FilterWire EZ increased the longitudinal vascular impedance the least after particles were injected (+29%). FiberNet increased longitudinal vascular impedance the most (+84%). CONCLUSION: This investigation, unique for examining the effects of physiologically realistic pulsatile flow on DPF performance, can aid in the development of future generations of novel DPFs. |
| | |
Authors:
|
Gail M Siewiorek; Mark H Wholey; Ender A Finol |
Publication Detail:
|
Type: Evaluation Studies; Journal Article; Research Support, N.I.H., Extramural; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists Volume: 16 ISSN: 1545-1550 ISO Abbreviation: J. Endovasc. Ther. Publication Date: 2009 Dec |
Date Detail:
|
Created Date: 2009-12-09 Completed Date: 2010-02-22 Revised Date: 2010-09-28 |
Medline Journal Info:
|
Nlm Unique ID: 100896915 Medline TA: J Endovasc Ther Country: United States |
Other Details:
|
Languages: eng Pagination: 735-43 Citation Subset: IM |
Affiliation:
|
Biomedical Engineering Department, Carnegie Mellon University, Pittsburgh, Pennsylvania 15213, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Angioplasty
/
adverse effects,
instrumentation* Carotid Arteries / pathology, physiopathology* Carotid Artery Diseases / pathology, physiopathology, therapy* Embolism / etiology, physiopathology, prevention & control* Equipment Design Filtration / instrumentation* Humans Materials Testing Microspheres Models, Anatomic Particle Size Pulsatile Flow* Stents* Time Factors |
| Comments/Corrections | |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Flow reversal for proximal neuroprotection during endovascular management of critical symptomatic ca...