In vitro performance assessment of distal protection devices for carotid artery stenting: effect of physiological anatomy on vascular resistance.
MedLine Citation:
PMID:  17924739     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To assess in vitro the performance of 5 distal protection devices (DPDs) by evaluating the capture efficiency, pressure gradient, volume flow rate, and vascular resistance in the internal carotid artery (ICA). METHODS: The time-averaged mean peak velocity in the common carotid artery and a blood-mimicking solution were used to simulate physiologICAl conditions in a silicone carotid phantom representing average human carotid artery geometry with a 70% symmetrICAl ICA stenosis. Five milligrams of dyed 200-microm nominal diameter polymer microspheres (larger than the pore size of the devices, except Spider RX, which was tested with 300-microm-diameter particles) were injected into the ICA. The percentages of particles missed after injection and lost during device retrieval were measured for the 5 devices (Spider RX, FilterWire EZ, RX Accunet, Angioguard XP, and Emboshield). The normalized pressure gradient, fraction of the volume flow rate, and vascular resistance in the ICA were calculated. RESULTS: Spider RX captured the most particles (missing 0.06%, p<0.05) and yielded the smallest normalized pressure gradient increase (4.2%), the largest volume flow rate fraction (0.40), and the smallest vascular resistance in the ICA (272 mmHg/L x min(-1), a 5.4% increase with respect to initial conditions). Angioguard XP captured the fewest particles (missing 36.3%, p<0.05 except Emboshield) and resulted in the largest normalized pressure gradient increase (37%) in the ICA. RX Accunet produced the smallest volume flow rate fraction in the ICA (0.30) and the largest vascular resistance in the ICA (470 mmHg/ L x min(-1), an 82.2% increase). Emboshield migrated approximately 6 cm distal to the original position after particle injection. FilterWire EZ lost the fewest particles during retrieval (0.45%, p<0.05 except Accunet RX and Spider RX) and had the best overall performance with 200-microm emboli (p<0.05 except Accunet RX). CONCLUSION: None of the devices tested completely prevented embolization. Overall, Spider RX had the best performance and is conjectured to have the best wall apposition of the devices tested. Vascular resistance should be considered a key filter design parameter for performance testing since it represents a quantitative estimation of the "slow-flow phenomenon." Our findings should be extrapolated cautiously to help interventionists choose the best device.
Authors:
Gail M Siewiorek; Mark H Wholey; Ender A Finol
Related Documents :
6759039 - Automated control of physiological variables and clinical therapy.
138669 - Turbulent airflow meter for long-term monitoring in patient-ventilator circuits.
16420959 - A field investigation of intra-cervical insemination with reduced sperm numbers in gilts.
7641339 - Blood pressure measurement in childhood epidemiological studies.
533759 - Effects of gas density on the frequency response of gas-filled pressure transducers.
15888859 - Performance characteristics of 10 home mechanical ventilators in pressure-support mode:...
18700629 - Detection of individuals prone to develop hypertension in their future life.
920819 - Physiologic determinants of left ventricular strength-interval curve of the dog.
16950369 - Electrical pacing counteracts intrinsic shortening of action potential duration of neon...
Publication Detail:
Type:  Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists     Volume:  14     ISSN:  1526-6028     ISO Abbreviation:  J. Endovasc. Ther.     Publication Date:  2007 Oct 
Date Detail:
Created Date:  2007-10-10     Completed Date:  2008-01-10     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  100896915     Medline TA:  J Endovasc Ther     Country:  United States    
Other Details:
Languages:  eng     Pagination:  712-24     Citation Subset:  IM    
Affiliation:
Biomedical Engineering Department, Carnegie Mellon University, Pittsburgh, Pennsylvania 15213, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Angioplasty / adverse effects,  instrumentation*
Blood Flow Velocity
Blood Pressure
Carotid Artery, Internal / pathology,  physiopathology,  surgery*
Carotid Stenosis / pathology,  physiopathology,  surgery*
Device Removal
Embolism / etiology,  pathology,  physiopathology,  prevention & control*
Equipment Design
Filtration / instrumentation*
Humans
Materials Testing
Models, Cardiovascular
Research Design
Stents*
Time Factors
Vascular Resistance*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  The role of carotid stenting.
Next Document:  A biodegradable stent based on poly(L-lactide) and poly(4-hydroxybutyrate) for peripheral vascular a...