Document Detail


In-vitro comparison of bag-valve-mask and the manually triggered oxygen-powered breathing device.
MedLine Citation:
PMID:  7621149     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To determine whether tidal volume, intrapleural pressure, and gastric volume differ when the bag--valve-- mask (BVM) and the manually triggered oxygen powered breathing device (MTBD) are used in the settings of normal and decreased lung compliance. METHODS: Prospective, randomized in-vitro model, with emergency medical technician-paramedics (EMT-Ps; n = 10) and emergency medical technicians (EMTs; n = 20) blinded to model gauges. Each participant randomly performed four one-minute trials ventilating a mechanical test lung through an intubation mannikin head at a rate of 20 breaths/min. Each subject used the BVM and the MTBD to ventilate the lung for 20 breaths during normal (0.10 L/cm H2O) and decreased 0.04 L/cm H2O) compliance. Tidal volumes and intrapleural pressures were recorded for every breath and the gastric volume was recorded for each trial using a flowmeter placed below a simulated esophageal sphincter (20 cm H2O PEEP valve). RESULTS: With normal compliance, mean tidal volumes and intrapleural pressures were not significantly different, while mean gastric volumes differed significantly (BVM = 1,300 mL, MTBD = 0 mL; p < 0.001). When compliance was decreased, mean tidal volumes and intrapleural pressures again did not differ significantly, while gastric volumes remained significantly different (BVM = 3,700 mL, MTBD = 1,100 mL p < 0.001). CONCLUSIONS: In this model, tidal volumes and intrapleural pressures were similar for the two devices. When compliance was normal, no participant insufflated the stomach with the MTBD, while the gastric volume with BVM ventilation averaged 1.3 L. With decreased compliance, the MTBD again delivered significantly less gastric volume than the BVM (1.1 vs. 3.7 L, respectively). These findings favoring MTBD ventilation require corroboration in vivo.
Authors:
J J Menegazzi; H J Winslow
Related Documents :
22007729 - Ocular pulse amplitude before and after cataract surgery.
22047619 - High frequency of diastolic dysfunction in a population-based cohort of elderly women--...
22245299 - Pressure transmission through ureteric stents: a novel in vivo human study.
21607169 - Non-dipper pattern is a determinant of the inappropriateness of left ventricular mass i...
22002029 - Markers of redox imbalance in the blood of hypertensive patients of a community in nort...
21612949 - Compression stockings with a negative pressure gradient have a more pronounced effect o...
15465369 - Venous hypertension, inflammation and valve remodeling.
23287369 - Hypertension and vascular dynamics in men and women with metabolic syndrome.
9707199 - Collagen organization in the branching region of human brain arteries.
Publication Detail:
Type:  Clinical Trial; Comparative Study; Journal Article; Randomized Controlled Trial    
Journal Detail:
Title:  Academic emergency medicine : official journal of the Society for Academic Emergency Medicine     Volume:  1     ISSN:  1069-6563     ISO Abbreviation:  Acad Emerg Med     Publication Date:    1994 Jan-Feb
Date Detail:
Created Date:  1995-08-28     Completed Date:  1995-08-28     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9418450     Medline TA:  Acad Emerg Med     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  29-33     Citation Subset:  IM    
Affiliation:
Center for Emergency Medicine of Western Pennsylvania, Pittsburgh 15213, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Humans
Lung Compliance
Pleura / physiology
Pressure
Prospective Studies
Respiration, Artificial / instrumentation*
Stomach
Tidal Volume