In-line moisture measurement during granulation with a four-wavelength near infrared sensor: an evaluation of particle size and binder effects.
MedLine Citation:
PMID:  10962238     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Factors affecting in-line near infrared (NIR) moisture measurement with a four-wavelength sensor were evaluated (choice of binder used in granulation liquid and the increase in particle size). An entire NIR spectrum is not necessary for the measurement of water, and often the use of only a few NIR wavelengths around the water band enables reliable and high-speed detection of moisture. Glass ballotini and microcrystalline cellulose (MCC) were used as model test materials. The binders studied were poly[1-(2-oxo-1-pyrrolidinyl)ethylene] (PVP) and gelatin. Full off-line NIR spectra of test materials at different levels of binder solutions were measured. The major spectral features for both the binders were bands around 1700 nm (first overtones CH related stretches) and 2200 nm (combination bands). Gelatin also had an NH band around 1500 nm (first overtones of NH stretches) and combination bands at about 2050 nm. Particle size effects were observed as an increase in spectra baseline. All these factors should be considered when choosing NIR wavelengths used for detection of water with a fixed wavelength set-up. A robust calibration model enables the development of in-process control of wet granulation processes.
Authors:
J Rantanen; E Räsänen; J Tenhunen; M Känsäkoski; J Mannermaa; J Yliruusi
Related Documents :
22454978 - Influence of some groundwater and surface waters constituents on the degradation of 4-c...
23253938 - Relative source contributions for perchlorate exposures in a lactating human cohort.
23241518 - Catalytic ozonation of phenol using copper coated pumice and zeolite as catalysts.
19463948 - Excipient-free nanoporous microparticles of budesonide for pulmonary delivery.
15664608 - Characterization of adsorbent composition in co-removal of hexavalent chromium with cop...
3252168 - Ultrastructural changes of synaptic glomeruli following intrathecal injection of distil...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  European journal of pharmaceutics and biopharmaceutics : official journal of Arbeitsgemeinschaft für Pharmazeutische Verfahrenstechnik e.V     Volume:  50     ISSN:  0939-6411     ISO Abbreviation:  Eur J Pharm Biopharm     Publication Date:  2000 Sep 
Date Detail:
Created Date:  2000-09-28     Completed Date:  2000-09-28     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9109778     Medline TA:  Eur J Pharm Biopharm     Country:  NETHERLANDS    
Other Details:
Languages:  eng     Pagination:  271-6     Citation Subset:  IM    
Affiliation:
Department of Pharmacy, Pharmaceutical Technology Division, University of Helsinki, Helsinki, Finland. jukka.rantanen@helsinki.fi
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Calibration
Particle Size
Powders
Spectroscopy, Near-Infrared
Technology, Pharmaceutical*