Improving alarm performance in the medical intensive care unit using delays and clinical context.
MedLine Citation:
PMID:  19372334     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
INTRODUCTION: In an intensive care unit, alarms are used to call attention to a patient, to alert a change in the patient's physiology, or to warn of a failure in a medical device; however, up to 94% of the alarms are false. Our purpose in this study was to identify a means of reducing the number of false alarms. METHODS: An observer recorded time-stamped information of alarms and the presence of health care team members in the patient room; each alarm response was classified as effective (action taken within 5 min), ineffective (no response to the alarm), and ignored (alarm consciously ignored or actively silenced). RESULTS: During the 200-h study period, 1271 separate entries by an individual to the room being observed were recorded, 1214 alarms occurred and 2344 tasks were performed. On average, alarms occurred 6.07 times per hour and were active for 3.28 min per hour; 23% were effective, 36% were ineffective, and 41% were ignored. The median alarm duration was 17 s. A 14-s delay before alarm presentation would remove 50% of the ignored and ineffective alarms, and a 19-s delay would remove 67%. Suctioning, washing, repositioning, and oral care caused 152 ignored or ineffective ventilator alarms. DISCUSSION: Introducing a 19-s alarm delay and automatically detecting suctioning, repositioning, oral care, and washing could reduce the number of ineffective and ignored alarms from 934 to 274. More reliable alarms could elicit more timely response, reduce workload, reduce noise pollution, and potentially improve patient safety.
Authors:
Matthias Görges; Boaz A Markewitz; Dwayne R Westenskow
Related Documents :
11487184 - Neurophysiological classification of ulnar entrapment across the elbow.
12003484 - Comparison of the safety and efficacy of an oscillating/rotating battery-powered toothb...
17161744 - The effect of random voice hand hygiene messages delivered by medical, nursing, and inf...
9121174 - Physical examination during space flight.
8901324 - Lungs at home.
10523444 - Nursing management of chronic wounds: best practices across the continuum of care.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Anesthesia and analgesia     Volume:  108     ISSN:  1526-7598     ISO Abbreviation:  Anesth. Analg.     Publication Date:  2009 May 
Date Detail:
Created Date:  2009-04-17     Completed Date:  2009-04-30     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1310650     Medline TA:  Anesth Analg     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1546-52     Citation Subset:  AIM; IM    
Affiliation:
Department of Anesthesiology, University of Utah, Salt Lake City, UT 84132, USA. matthias@abl.med.utah.ed
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Artifacts
Attention
Automation
Equipment Design
Equipment Failure*
Female
Humans
Intensive Care / methods*
Intensive Care Units*
Male
Middle Aged
Monitoring, Physiologic / instrumentation*
Noise / prevention & control
Safety Management
Time Factors
Ventilators, Mechanical
Workload
Comments/Corrections
Comment In:
Anesth Analg. 2009 May;108(5):1382-3   [PMID:  19372311 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Isovaleric, methylmalonic, and propionic acid decrease anesthetic EC50 in tadpoles, modulate glycine...
Next Document:  Noninvasive assessment of cardiac index in healthy volunteers: a comparison between thoracic impedan...