Document Detail


Improvement in the accuracy of dual energy x-ray absorptiometry for whole body and regional analysis of body composition: validation using piglets and methodologic considerations in infants.
MedLine Citation:
PMID:  9098865     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Previously, we conducted dual energy x-ray absorptiometry (DXA) (Hologic QDR-1000/W) scans and carcass analysis of piglets to evaluate the Pediatric Whole Body software (PedWB) (V5.35) for use in infants. A software upgrade designed for infant whole body (InfWB) (V5.56) led to a reassessment of DXA by: 1) reanalysis of the original scans using InfWB software and 2) comparison of InfWB-estimates of bone mineral content (BMC) and lean and fat mass with chemical analysis. Other assessments included 1) methods of regional analysis and 2) artifacts and the Infant Table Pad in the scan field. The mean coefficients of variation for InfWB whole body measures in small piglets (n = 10, weight 1575 +/- 73 g) and large piglets (n = 10, weight 5894 +/- 208 g) were less than 2.6% except for fat mass which was higher (8.0% versus 6.3% and 6.6% versus 3.5%, respectively) compared with PedWB. In large piglets InfWB produced good estimates of BMC, lean and fat masses. In small piglets, fat mass by InfWB was correlated with chemical analysis, but not by PedWB. There was improvement in the estimation of BMC with InfWB, from 27 +/- 2.2 g to 32 +/- 2.3 g (carcass ash = 38 +/- 3.3 g). Femur BMC analysis by InfWB was precise and was accurate when compared with chemical analysis. Artifacts in the DXA scan field (diapers and blankets) resulted in an increase of the DXA-estimated fat and lean masses. The Infant Table Pad increased the estimate of fat mass in a small piglet by 50%, thus further study is required before it is used routinely. Improvements of the DXA technology have resulted in a more accurate tool, if scanning procedures are carefully implemented.
Authors:
J A Brunton; H A Weiler; S A Atkinson
Related Documents :
10147455 - Near-infrared interactance (nir): a new non-invasive technique to estimate subcutaneous...
16553245 - Water and heat--the priority for the newborn infant.
11483805 - A clinical comparison of radiant warmer and incubator care for preterm infants from bir...
10905455 - Reactivity of a wollastonite-tricalcium phosphate bioeutectic ceramic in human parotid ...
12777545 - Infants with bilirubin levels of 30 mg/dl or more in a large managed care organization.
6696545 - Neonatal pneumopericardium: a surgical emergency.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Pediatric research     Volume:  41     ISSN:  0031-3998     ISO Abbreviation:  Pediatr. Res.     Publication Date:  1997 Apr 
Date Detail:
Created Date:  1997-07-02     Completed Date:  1997-07-02     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  0100714     Medline TA:  Pediatr Res     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  590-6     Citation Subset:  IM    
Affiliation:
Department of Pediatrics, McMaster University, Hamilton, Ontario, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Absorptiometry, Photon*
Animals
Artifacts
Body Composition*
Female
Humans
Infant, Newborn
Infant, Premature*
Linear Models
Reproducibility of Results
Swine

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Energy expended by low birth weight infants in the deposition of protein and fat.
Next Document:  The oral absorption of micro- and nanoparticulates: neither exceptional nor unusual.