Home blood pressure measurement: validation of the Braun BP 2550 (UG) monitor according to the ESH International Protocol.
MedLine Citation:
PMID:  15021079     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The Sixth Report of the Joint National Committee states that the 'measurement of blood pressure outside the clinician's office may provide valuable information for the initial evaluation of patients with hypertension and for monitoring the response to treatment'. Prior to their marketing, however, blood pressure measuring devices should be independently validated for safety and accuracy using standard guidelines. OBJECTIVE: The main objective of the present study was the clinical evaluation of the Braun Precision Sensor 2550 upgrade (Braun BP 2550 (UG)) according to the International Protocol of the European Society of Hypertension (ESH). METHODS: Measurements of systolic and diastolic blood pressure in 33 subjects were obtained using the mercury sphygmomanometer (Baumanometer) and the device. The sizes of the cuffs used for different arm circumferences were 12 x 23 cm, 15 x 33 cm and 18 x 36 cm for the thin adult, adult and overweight adult, respectively. The adult size was mostly used. If the cuff of the sphygmomanometer did not fall between the two range lines, a larger or smaller cuff was required. The reproducibility of self-blood pressure measurement was compared in 33 subjects (99 measurements). Nine sequential same-arm measurements comparing the test device and the standard mercury sphygmomanometer were carried out according to the International Protocol of the ESH in three steps (phases 1, 2.1 and 2.2). RESULTS: The device passed phases 1 and 2.1 as 75 systolic and 78 diastolic readings fell within 5 mmHg, exceeding the total of 65 required for approval. It also passed phase 2.2 as 30 subjects for systolic and 29 for diastolic had at least two out of three comparisons within 5 mmHg, exceeding the 22 subjects required. The difference-against-mean plots showed that the device-sphygmomanometer disagreement was -1.58+/-4.84 for systolic pressure (the Braun being lower) and +2.23+/-3.84 for diastolic pressure (the Braun being higher), respectively. CONCLUSIONS: The equipment passed both phases of the International Protocol (subject measurements and subject accuracy). The difference of 0-5 mmHg between the sphygmomanometer and the device represents 'measurements considered to be very accurate and with no error of clinical relevance'. It can therefore be used at home as an adjunct to physicians' effort to control blood pressure in their hypertensive patients.
Authors:
Hector Nolly; Marcelo Romero; Alejandro Nolly; Pablo Osso; Omar Reinoso; Mariela Nolly
Related Documents :
7425819 - Pressure measurements: device for obtaining readouts from multiple sites using pneumati...
2627409 - Anesthesia during total electrical failure, or what would you do if the lights went out?
9246869 - An experimental high energy therapeutic ultrasound equipment: design and characterisation.
10362889 - Evaluation of the schiller br-102 ambulatory blood pressure system according to the pro...
10386739 - Intercomparison of acoustic output measurements of a diagnostic ultrasound device.
12870259 - Advances in icp monitoring techniques.
17541209 - Reduction of qtc dispersion by the angiotensin ii receptor blocker valsartan may be rel...
14687169 - A pilot study to investigate the use of oxpentifylline (pentoxifylline) and thalidomide...
3083629 - Pulmonary blood flow reduction by prostaglandin f2 alpha and pulmonary artery balloon m...
Publication Detail:
Type:  Comparative Study; Journal Article; Validation Studies    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  9     ISSN:  1359-5237     ISO Abbreviation:  Blood Press Monit     Publication Date:  2004 Feb 
Date Detail:
Created Date:  2004-03-15     Completed Date:  2004-10-19     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  53-8     Citation Subset:  IM    
Affiliation:
Hypertension Center, Mendoza, Argentina.nolly@supernet.com.ar
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Anthropometry
Arm
Blood Pressure
Blood Pressure Determination / instrumentation*,  standards
Blood Pressure Monitors / standards*
Guideline Adherence
Home Care Services / standards*
Humans
Reproducibility of Results
Sphygmomanometers / standards

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Overall accuracy of the BpTRU--an automated electronic blood pressure device.
Next Document:  Debridement: controlling the necrotic/cellular burden.