High revision rate of hydroxyapatite-coated ABG-I prosthesis.
MedLine Citation:
PMID:  19802665     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The aim of this study was to investigate the causes of the high revision rate of the hydroxyapatite-coated Anatomique Benoist Girard (ABG-I) prosthesis. METHODS: We performed 204 total hip arthroplasties (THAs) at our hospital between March 1992 and December 1996. Of the 204 THA patients, 129 were followed up; the mean duration of follow-up was 12.2 years (range 10.0-14.5 years). There were 113 male patients and 16 female patients, with an average age of 53 years (30-83 years) at the time of surgery. The reasons for THA were avascular necrosis in 119 hips, acetabular dysplasia in 8 hips, and traumatic arthritis in 2 hips. RESULTS: The Harris hip score was 47.3 preoperatively and 86.4 at the last follow-up. The linear polyethylene wear was an average of 0.29 mm/year. Acetabular osteolysis was seen in 113 cases (88%). Altogether, 61 (47.2%) acetabular cups were revised for aseptic loosening in 38 hips, polyethylene wear and osteolysis in 20 hips, recurrent dislocation in 2 hips, and deep infection in 1 hip. Femoral osteolysis was observed in 100 cases (77.5%). The femoral stem was revised in 4 hips (3.1%). CONCLUSION: We observed that the fatal detriment to ABG-I acetabular cup survival in long-term follow-up was periacetabular osteolysis followed by aseptic loosening.
Authors:
Young-Yool Chung; Sung-Chan Ki; Kwang-Young So; Dae-Hee Kim; Kyung Ho Park; Yeon Soo Lee
Related Documents :
17562405 - A randomized clinical trial of cementless femoral stems with and without hydroxyapatite...
11284595 - Long-term results after operative treatment of femoral neck fractures with ceramic head...
9699815 - Comparison of the wrightington fc hip with the charnley low-friction arthroplasty. 10- ...
18555645 - Acute liner disassociation of a pinnacle acetabular component.
8362425 - Value of single-photon emission-computed tomography in acute stroke therapeutic trials.
15060765 - Prevention of intravascular catheter-related infection with newer chlorhexidine-silver ...
Publication Detail:
Type:  Journal Article     Date:  2009-10-03
Journal Detail:
Title:  Journal of orthopaedic science : official journal of the Japanese Orthopaedic Association     Volume:  14     ISSN:  1436-2023     ISO Abbreviation:  J Orthop Sci     Publication Date:  2009 Sep 
Date Detail:
Created Date:  2009-10-05     Completed Date:  2010-01-14     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9604934     Medline TA:  J Orthop Sci     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  543-7     Citation Subset:  IM    
Affiliation:
Department of Orthopedic Surgery, Gwangju Christian Hospital, Gwangju, Republic of Korea.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Arthroplasty, Replacement, Hip / adverse effects,  instrumentation*
Durapatite*
Equipment Failure Analysis*
Female
Follow-Up Studies
Hip Prosthesis*
Humans
Kaplan-Meiers Estimate
Male
Middle Aged
Prosthesis Failure*
Reoperation