| Hemodynamic energy delivery of the pulsatile flow in a simulated pediatric extracorporeal circuit. | |
| | |
MedLine Citation:
|
PMID: 19092666 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Objectives of this study were 1) to precisely measure the generation and delivery of hemodynamic energy levels during pediatric cardiopulmonary bypass procedure using a simulated circuit and 2) to quantify energy losses of the extracorporeal circuit components. A conventional roller pump, pediatric hollow-fiber membrane oxygenator, arterial filter, arterial cannula, and (1/4)-in tubing were used. The circuit was primed with 40/60 glycerin/water mixture. Postcannula (pseudo patient's pressure) pressure was maintained 40 mm Hg by a Hoffman clamp. All trials were conducted at 800 mL/min of pump flow with pulsatile (n=6) and nonpulsatile (n=6) modes at room temperature. Simultaneous blood flow and pressures at pre/post oxygenator and pre/post cannula sites were recorded. There were gradual decreases in values of circuit mean pressure, energy equivalent pressure, surplus hemodynamic energy, and total hemodynamic energy (THE) from pre- and postoxygenator to pre- and postcannula sites. Approximately 30% of THE generated by roller pump was damped by the membrane oxygenator. About 25% of THE was lost by the arterial filter and the length of the arterial tubing. Only 15%-18% of THE was delivered into the pseudo patient. Although the results of this study showed that the roller pump could provide higher hemodynamic energy per second with pulsatile flow compared with the nonpulsatile flow, the majority (80%) of hemodynamic energy was damped by the components of the extracorporeal circuit. |
| | |
Authors:
|
Shigang Wang; Nikkole Haines; Akif Undar |
Related Documents
:
|
2261016 - Left ventricular assistance with the centrifugal pump: management of the patient with s... 18001396 - Continuous renal replacement therapy with a pulsatile tubular pump: parameters must be ... 8573846 - Acute phase responses of vasoactive hormones to non pulsatile systemic circulation. 16018466 - An aeroacoustically driven thermoacoustic heat pump. 8656626 - Inhibition of no synthase prevents acute collateral artery dilation in the rat hindlimb. 3707786 - Ambulatory blood pressure and assessment of pacemaker function. |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: ASAIO journal (American Society for Artificial Internal Organs : 1992) Volume: 55 ISSN: 1538-943X ISO Abbreviation: ASAIO J. Publication Date: 2009 Jan-Feb |
Date Detail:
|
Created Date: 2009-01-13 Completed Date: 2009-03-19 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 9204109 Medline TA: ASAIO J Country: United States |
Other Details:
|
Languages: eng Pagination: 96-9 Citation Subset: IM |
Affiliation:
|
Pediatric Cardiac Research Laboratory, Department of Pediatrics, Penn State Milton S. Hershey Medical Center, Penn State College Medicine, Penn State Hershey Children's Hospital, Hershey, Pennsylvania 17033-0850, USA. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Cardiopulmonary Bypass
/
instrumentation,
methods Child Extracorporeal Membrane Oxygenation* Heart-Lung Machine Humans Models, Cardiovascular* Pulsatile Flow / physiology* |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Artificial muscle to wash blood out of fibrillating atrium: an alternative to lifelong anticoagulati...
Next Document: In vivo experimental testing of the FW axial blood pump for left ventricular support in Fu Wai Hospi...