Document Detail


Headform and n95 filtering facepiece respirator interaction: contact pressure simulation and validation.
MedLine Citation:
PMID:  22168255     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
This article presents a computational and experimental study of contact pressure between six N95 filtering facepiece respirators (FFRs) and five newly developed digital headforms (small, medium, large, long/narrow, and short/wide). Contact interaction is simulated using the finite element method and validated by experiments using a pressure mapping system. The headform model has multiple layers: a skin layer, muscle layer, fatty tissue layer, and bone layer. Each headform is divided into five parts (two parts for the cheeks, one part for the upper forehead, one part for the chin, and one part for the back side of the head). Each respirator model comprises multiple layers and two straps. The simulation process has two stages for each respirator/headform combination. The first stage is to wrap the straps around the back of the headform and pull the respirator away from the face. The second stage is to release the respirator so that the respirator moves toward the face. Strap forces and contact interactions are generated between the respirators and the headforms. Meanwhile, a real-time surface pressure mapping system is used to record the pressures at six key locations to validate the computational results. There is a strong correlation between computational and experimental results (R(2) = 0.88). By comparing the pressure values from simulations and experiments, we have validated the simulation models.
Authors:
Zhipeng Lei; Jingzhou James Yang; Ziqing Zhuang
Related Documents :
20379045 - End-tidal carbon dioxide concentration can estimate the appropriate timing for weaning ...
1648015 - The hypoxic stress on erythrocytes associated with superoxide formation.
22225415 - Determination of expander apparatus displacements and contact pressures on the mucosa u...
8685755 - Improvement in oxygenation after large volume paracentesis.
6464055 - Naloxone ameliorates the pathophysiologic changes which lead to and attend an acute str...
7605055 - Dynamics of a mechanical monoleaflet heart valve prosthesis in the closing phase: effec...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Journal of occupational and environmental hygiene     Volume:  9     ISSN:  1545-9632     ISO Abbreviation:  J Occup Environ Hyg     Publication Date:  2012 Jan 
Date Detail:
Created Date:  2011-12-15     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101189458     Medline TA:  J Occup Environ Hyg     Country:  England    
Other Details:
Languages:  eng     Pagination:  46-58     Citation Subset:  IM    
Affiliation:
a Human-Centric Design Research Lab, Department of Mechanical Engineering , Texas Tech University , Lubbock , Texas.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Assessing potential nanoparticle release during nanocomposite shredding using direct-reading instrum...
Next Document:  Discomfort and exertion associated with prolonged wear of respiratory protection in a health care se...