Document Detail


Hazard report. Smiths Medical Medfusion 3010a syringe pump may overinfuse if software is outdated.
MedLine Citation:
PMID:  21305937     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
Smiths Medical Medfusion 3010a syringe pumps that aren't equipped with the latest software version (2.0.6) could overinfuse if the "recall last settings" function is used following an infusion in volume/time mode. If your facility has pumps equipped with software versions older than 2.0.6, contact Smiths Medical to obtain an upgrade.
Authors:
-
Related Documents :
20428307 - How "objective" are psychiatric diagnoses?: (guess again).
2480937 - Medicine as metaphor and medium in group psychotherapy with psychiatric patients.
2585427 - Prevalence, detection and referral of psychiatric morbidity in general medical patients.
19174297 - Self-reported health problems among travelers visiting cuzco: a peruvian airport survey.
20375507 - Mild cognitive impairment in medical inpatients: the mini-mental state examination is a...
537527 - A place for oncology in medical undergraduate teaching.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Health devices     Volume:  39     ISSN:  0046-7022     ISO Abbreviation:  Health Devices     Publication Date:  2010 Apr 
Date Detail:
Created Date:  2011-02-10     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1262063     Medline TA:  Health Devices     Country:  United States    
Other Details:
Languages:  eng     Pagination:  137-8     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Hazard report. Arthrex AR-6400 and AR-6475 arthroscopy pumps can deliver excessive pressure without ...
Next Document:  Hazard report. Parks Medical Flo-Lab 2100-SX may interpret IR emissions from CenTrak RTLS as user co...