Document Detail


Hazard report. Arthrex AR-6400 and AR-6475 arthroscopy pumps can deliver excessive pressure without alerting user.
MedLine Citation:
PMID:  21305936     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
Under certain conditions, Arthrex AR-6400 and AR-6475 arthroscopic irrigation/distention pumps can create pressures within the patient's joint that are more than twice the desired pressure without alerting the clinician, increasing the potential for serious patient injury due to extracapsular solution infiltration. Facilities need to thoroughly understand the behavior of these pumps and determine whether the clinical benefits of using them outweigh the solution-infiltration concerns. If your facility is not comfortable with the behavior of these pumps, use an alternative pump.
Authors:
-
Related Documents :
2472846 - Hypoxia- and endothelium-mediated changes in the pharmacological responsiveness of circ...
6777846 - Comparison of steady state pulmonary diffusing capacity estimates for o2 and co in dogs.
11507176 - Effect of hypoxia on the hypopnoeic and apnoeic threshold for co(2) in sleeping humans.
8252896 - Metabolic acidemia with hypoxia attenuates the hemodynamic responses to epinephrine dur...
11130496 - Effect of web 2086 on leukocyte adherence in response to hemorrhagic shock in rats.
9444886 - Antihypertensive agents that limit ventricular hypertrophy inhibit cardiac expression o...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Health devices     Volume:  39     ISSN:  0046-7022     ISO Abbreviation:  Health Devices     Publication Date:  2010 Apr 
Date Detail:
Created Date:  2011-02-10     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1262063     Medline TA:  Health Devices     Country:  United States    
Other Details:
Languages:  eng     Pagination:  134-6     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Consensus in rhinoplasty.
Next Document:  Hazard report. Smiths Medical Medfusion 3010a syringe pump may overinfuse if software is outdated.