Document Detail


Graft flow increases with release of stabilizing device in off-pump coronary surgery.
MedLine Citation:
PMID:  15003100     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Transit-time Doppler can be used for intraoperative assessment of graft patency, by measuring conduit flow, or from the calculation of a derivation of the spectral waveform. The optimal timing of measurement is when graft flows are maximal. Fifteen patients underwent multivessel coronary bypass surgery without the use of cardiopulmonary bypass. Transit-time Doppler of 57 anastomoses was performed immediately before and after release of the Octopus stabilizing device, and mean graft flow was measured as over 5 consecutive beats. The graft flow increased after release of the stabilizing device for all grafts from 20.1+/-1.8 to 33.2+/-2.4 ml/min (P<0.001). When analyzed by individual grafts, and by coronary artery territory, the increase in flow remained significant. The timing of transit-time Doppler assessment of conduit flow should be performed immediately after release of the stabilizing device.
Authors:
Alistair G Royse; Colin F Royse; Christina T Wong
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia     Volume:  9     ISSN:  1341-1098     ISO Abbreviation:  Ann Thorac Cardiovasc Surg     Publication Date:  2003 Dec 
Date Detail:
Created Date:  2004-03-08     Completed Date:  2004-04-15     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9703158     Medline TA:  Ann Thorac Cardiovasc Surg     Country:  Japan    
Other Details:
Languages:  eng     Pagination:  384-8     Citation Subset:  IM    
Affiliation:
Department of Pharmacology, The University of Melbourne, Victoria, Australia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Anastomosis, Surgical / instrumentation*
Coronary Artery Bypass*
Coronary Circulation*
Echocardiography, Doppler
Female
Humans
Male
Middle Aged
Pressure
Suction
Treatment Outcome
Vascular Patency*