Document Detail


Follow-up assessment of St. Jude Medical prosthetic valve in the tricuspid position: clinical and hemodynamic results.
MedLine Citation:
PMID:  6712333     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Seven patients underwent postoperative right heart catheterization implantation of the St. Jude Medical prosthetic valve in the tricuspid position. Six patients were in atrial fibrillation at catheterization, and 1 was in normal sinus rhythm. At postoperative catheterization, the mean right atrial pressure ranged between 4 and 16 mm Hg (mean, 9.7 mm Hg); right ventricular systolic pressure was normal in 1 patient, mildly elevated (less than 50 mm Hg) in 4 patients, and moderately elevated (65 and 70 mm Hg) in 2. The cardiac output ranged between 3.0 and 7.0 L/min (mean, 4.2 L/min). There was no end-diastolic gradient across the St. Jude Medical prosthesis in 6 patients. The other patient had a gradient of 2 mm Hg across the valve when cardiac output was 7.0 L/min. On fluoroscopy, both discs demonstrated full excursion in all patients. These data demonstrate that a normally functioning St. Jude Medical valve in the tricuspid position does not create obstruction to forward flow, and they support use of this prosthesis in patients with tricuspid valve disease.
Authors:
A K Singh; F D Christian; D O Williams; C S Georas; R R Riley; K B Nanian; K E Karlson
Related Documents :
16846423 - The "button inside" technique for the aortic root replacement: a modified button techni...
19185153 - Mechanical valves versus the ross procedure for aortic valve replacement in children: p...
6198743 - Haemostatic variables and the outcome of myocardial infarction.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  The Annals of thoracic surgery     Volume:  37     ISSN:  0003-4975     ISO Abbreviation:  Ann. Thorac. Surg.     Publication Date:  1984 Apr 
Date Detail:
Created Date:  1984-05-02     Completed Date:  1984-05-02     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  15030100R     Medline TA:  Ann Thorac Surg     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  324-7     Citation Subset:  AIM; IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Blood Pressure
Cardiac Output
Female
Follow-Up Studies
Heart Catheterization
Heart Rate
Heart Valve Prosthesis* / standards
Hemodynamics*
Humans
Male
Middle Aged
Postoperative Period
Tricuspid Valve / surgery

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine