Document Detail


Flow-dependent properties of positive expiratory pressure devices.
MedLine Citation:
PMID:  7613548     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Valves for positive expiratory pressure (PEP) can be characterized as threshold resistors, ideally providing pressure independent of the expiratory flow, or as flow-dependent resistors. The aim of the study was to evaluate the flow-dependence properties of PEP devices of the flow resistor type compared to threshold resistor devices. Pressures were measured on three different flow resistor valves: the PEP-mask, the Pari-PEP-System and the System 22-PEP with orifice diameters of 1.5-5.0 mm; and on three threshold resistors, the underwater seal, the Ambu Positive End-Expiratory Pressure (PEEP) valve and the Vital Signs PEEP valve with pressures of 0, 5, 10, 15 and 20 cmH2O. All devices were studied with constant flows of 3, 6, 9, 12, 15, 18, 30, 60, 90, 120 and 150 l.min-1. The PEP-mask, the Pari-PEP and the System 22-PEP showed the typical pattern of flow resistors, i.e. a pressure increasing with flow, dependent on the diameter of the orifice. The underwater seal and the Vital Signs PEEP valves acted as almost ideal threshold resistors. The Ambu PEEP valves acted as threshold resistors at the lower flows, but showed flow-dependency at higher flows. The Vital Signs PEEP valves gave lower pressures and Ambu PEEP valves gave higher pressures compared with indicated values, whereas the underwater seal gave the intended pressure. In clinical use of PEP treatment the actual pressure should be measured to ensure the intended pressure, no matter which type of resistor is used.
Authors:
E F Christensen; R H Jensen; N K Schønemann; K D Petersen
Related Documents :
17194318 - Setting positive end-expiratory pressure during jet ventilation to replicate the mean a...
20193478 - Alveolar stability under different combinations of positive end-expiratory pressure and...
11175238 - Lung collapse during low tidal volume ventilation in acute respiratory distress syndrome.
7036798 - Pulmonary oxygen exchange during endobronchial anesthesia: effect of tidal volume and p...
11225178 - Saponin content and trypsin inhibitor activity in processed and cooked pigeon pea culti...
3317338 - Comparative efficacy of captopril and atenolol in moderately severe essential hypertens...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Monaldi archives for chest disease = Archivio Monaldi per le malattie del torace / Fondazione clinica del lavoro, IRCCS [and] Istituto di clinica tisiologica e malattie apparato respiratorio, Università di Napoli, Secondo ateneo     Volume:  50     ISSN:  1122-0643     ISO Abbreviation:  Monaldi Arch Chest Dis     Publication Date:  1995 Apr 
Date Detail:
Created Date:  1995-08-18     Completed Date:  1995-08-18     Revised Date:  2008-06-02    
Medline Journal Info:
Nlm Unique ID:  9307314     Medline TA:  Monaldi Arch Chest Dis     Country:  ITALY    
Other Details:
Languages:  eng     Pagination:  150-3     Citation Subset:  IM    
Affiliation:
Dept of Anesthesiology, Arhus Kommunehospital, University Hospital of Arhus, Denmark.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Equipment Design
Humans
Materials Testing
Positive-Pressure Respiration / instrumentation*
Pressure
Rheology
Surface Properties

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Plethysmographic measurements in the clinical assessment of infants with bronchopulmonary disease.
Next Document:  Asthma and bronchial reactivity: a contribution from multivariate analysis.