Failures of epicardial and transvenous leads for implantable cardioverter defibrillators.
MedLine Citation:
PMID:  7484734     Owner:  NLM     Status:  MEDLINE    
Lead failures constitute a major risk for patients with an implantable cardioverter defibrillator (ICD). However, data about the incidence and patterns of ICD-lead failures in a larger population are lacking. We analyzed the short-term and midterm performance of 27 epicardial and 103 nonthoracotomy ICD-lead systems during a follow-up period of 36 +/- 21 months and 22 +/- 10 months, respectively (p < 0.05). The failure rate was 5 (19%) of 27 in the epicardial and 6 (6%) of 103 in the nonthoracotomy group (p < 0.05). The most common symptom was erroneous detection of ventricular fibrillation from artifact sensing in five patients. Two patients had to be resuscitated because of failure to defibrillate. Loss of pacing and loss of sensing were seen in two patients. Only two asymptomatic lead fractures could be diagnosed on routine radiograph. In conclusion, there was a considerable rate of lead failures, especially in epicardial systems. Long-term studies addressing the longevity of ICD leads, mechanisms of their failures, and improved diagnostic facilities are important to further increase the safety of this therapeutic approach.
S Mattke; D Müller; A Markewitz; H Kaulbach; M Schmöckel; U Dorwarth; E Hoffmann; G Steinbeck
Related Documents :
8751504 - Right ventricular ejection fraction in the preoperative risk evaluation of candidates f...
20837894 - Long-term benefit of primary prevention with an implantable cardioverter-defibrillator:...
15082274 - Early replacement of pulmonary valve after repair of tetralogy: is it really beneficial?
19497444 - Incidence, risk factors, and outcome of traumatic tricuspid regurgitation after percuta...
16987114 - A community model of group therapy for the older patients with chronic obstructive pulm...
23196094 - The intestinal absorption of orally administered hydrolytic enzymes and their effects i...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  American heart journal     Volume:  130     ISSN:  0002-8703     ISO Abbreviation:  Am. Heart J.     Publication Date:  1995 Nov 
Date Detail:
Created Date:  1995-12-21     Completed Date:  1995-12-21     Revised Date:  2006-02-27    
Medline Journal Info:
Nlm Unique ID:  0370465     Medline TA:  Am Heart J     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  1040-4     Citation Subset:  AIM; IM    
Department of Medicine I, University of Munich, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Defibrillators, Implantable*
Equipment Failure
Retrospective Studies

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  Prospective study of retrograde coronary venography in patients with posteroseptal and left-sided ac...
Next Document:  Asymmetric thickness of the left ventricular wall resulting from asynchronous electric activation: a...