Document Detail


Factors affecting the validity of the standard blood pressure cuff.
MedLine Citation:
PMID:  8119055     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Fifty-eight patients, 31 women and 27 men, with a wide range of upper arm circumferences (20-43 cm), ages (16-82 years) and body mass indices (16-55 kg m-2) were studied. The direct blood pressure was measured in the brachial artery in one arm and simultaneously the indirect pressure was measured in the other arm using a large standard cuff (rubber bag 12 x 35 cm). The mean intra-arterial blood pressure was (systolic/diastolic) 137 +/- 25 mmHg (SD)/76 +/- 10 mmHg. With the cuff the mean blood pressure was 134 +/- 23 mmHg/85 +/- 11 mmHg. The cuff significantly underestimated the systolic blood pressure (3.2 +/- 11.4 mmHg, range -25-30 mmHg) while the diastolic pressure was overestimated significantly (8.8 +/- 8.5 mmHg, range -32-14 mmHg). The direct/indirect blood pressure difference correlated significantly to the intra-arterial systolic and diastolic pressures (systolic/diastolic; r = 0.44, P = 0.0006/r = 0.36, P = 0.0062), but only to the auscultatory diastolic pressure (r = -0.02, P = 0.9/r = -0.45, P = 0.0004). The direct/indirect difference correlated significantly to the arm circumference (r = -0.33, P = 0.01/r = -0.30, P = 0.02) as did the diastolic direct/indirect difference to pulse rate (r = 0.34, P = 0.009). Age did not correlate significantly to the direct/indirect blood pressure difference in this study (r = -0.02, P = 0.9/r = -0.10, P = 0.5). It can be concluded that in addition to the ratio between the arm circumference and the width of the cuff (rubber bag), there are other sources of systematic errors of indirect blood pressure measurement, in particular pulse rate and the level of indirect blood pressure. If corrections could be made with regard to these variables, it is obvious that the reliability of the indirect method, especially when used in epidemiological contexts, would increase.
Authors:
M Stolt; G Sjönell; H Aström; L Hansson
Related Documents :
10226205 - Validation of cas model 9010 automated blood pressure monitor:children/adult and neonat...
17012005 - Interface pressure and stiffness of ready made compression stockings: comparison of in ...
11942865 - Accuracy of radial arterial pressure measurement during surgery under controlled hypote...
19308725 - The "weight" of words on the forearms during relaxation.
17091855 - A high-sensitivity, dual-plate, thickness-shear mode pressure sensor.
1537095 - Cryoblockade in limbic brain (amygdala) prevents or delays ventricular fibrillation aft...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Clinical physiology (Oxford, England)     Volume:  13     ISSN:  0144-5979     ISO Abbreviation:  Clin Physiol     Publication Date:  1993 Nov 
Date Detail:
Created Date:  1994-04-06     Completed Date:  1994-04-06     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  8309768     Medline TA:  Clin Physiol     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  611-20     Citation Subset:  IM    
Affiliation:
Kvartersakuten Birka, Stockholm, Sweden.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Aged
Aged, 80 and over
Aging / physiology
Arm / anatomy & histology,  physiology
Blood Pressure / physiology*
Blood Pressure Determination / instrumentation*
Diastole / physiology
Evaluation Studies as Topic
Female
Humans
Male
Middle Aged
Pulse / physiology
Reproducibility of Results
Systole / physiology

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Changes in skin blood flow in ischaemic limbs after vascular reconstruction measured by fluorescein ...
Next Document:  Postural effects on diastolic blood pressure are differently recorded by a non-invasive ambulatory b...