Extraction of the inner coil of a pacemaker lead slid into the pulmonary artery.
MedLine Citation:
PMID:  17338727     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A patient required lead extraction for chronic draining sinus, due to abandoned leads. Preoperatively, the chest film showed a filament in the right pulmonary artery: it was the inner coil of an old atrial lead that, while remaining anchored to the auricle, slid outside the outer coil. By right subclavian approach, the old ventricular lead and the outer coil of the atrial lead were removed. Then, by right jugular approach, the freely floating end of the inner atrial coil was grasped by a pig-tail catheter, drawn back into the superior vena cava, exteriorized by a Lasso catheter, and finally extracted.
Authors:
Pier Giorgio Golzio; Maria Grazia Bongiorni; Amedeo Chiribiri; Erica Franco
Related Documents :
17909777 - Endovascular treatment of basilar and ica termination aneurysms: effects of the use of ...
16950447 - Combined open and endovascular treatment of a saccular aneurysm and redundant loop of t...
17827897 - Evaluation of the efficacy and safety of coil occlusion for patent ductus arteriosus in...
20557757 - Treatment of arteriovenous malformations with hydrocoils in a swine model.
20140757 - A case of hughes-stovin syndrome associated with hyperhomocysteinemia.
18449427 - Clinical and echocardiographic measures governing thromboembolism destination in atrial...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  30     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2007 Feb 
Date Detail:
Created Date:  2007-03-06     Completed Date:  2007-05-01     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  273-5     Citation Subset:  IM    
Affiliation:
Cardiology Division, Department of Internal Medicine, University of Torino, Torino, Italy. pgolzio@molinette.piemonte.it
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Device Removal / methods*
Electrodes, Implanted / adverse effects*
Female
Foreign-Body Migration / etiology*,  surgery*
Heart Catheterization / instrumentation,  methods*
Humans
Pacemaker, Artificial / adverse effects*
Pulmonary Artery / surgery*
Treatment Outcome

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Using the upper limit of vulnerability to assess defibrillation efficacy at implantation of ICDs.
Next Document:  Successful placement of left ventricular pacing lead despite coronary sinus perforation into the per...