Extracorporeal bubbles: a word of caution.
MedLine Citation:
PMID:  20197351     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: Gaseous microemboli (GME) are a cause of neurocognitive deficit postcardiac surgery with cardiopulmonary bypass. However, the measurement of microbubbles during cardiopulmonary bypass is not easy, because blood is an opaque fluid and contains microparticles. Recently, two new microbubble counters, the Gampt BC200 and the emboli detection and classification (EDAC) quantifier, were developed for use during cardiac surgery. The accuracy of both devices was validated against industrial standards. METHODS: A clear blood analogue (30%/70% glycerol-water) was pumped, by means of a rollerpump out of a partially filled arterial line screen filter with a nominal pore size of 20 microm. Downstream the pump, all test sensors and an optical cuvette, were inserted in a vertically mounted 3/8 inch tubing. Measurements were taken at flows of 405, 1080, 3000 and 6000 ml/min. Backlight shadowgraphy and optical counting were used as industrial reference techniques for measuring size and counts. RESULTS: On average the EDAC quantifier underestimates average bubble diameter by 35+/-1%, 13+/-8%, 71+/-7% and 33+/-4% at 405, 1080, 3000 and 6000 ml/min, respectively. The Gampt BC200 has good sizing agreement at 1080 ml/min (+5+/-17%) and 3000 ml/min (+33+/-6%) but overestimates average diameter by 220+/-40% and 295+/-49% at 405 and 6000 ml/min, respectively. Both devices only partially count the number of microbubbles at higher flows. At 3 l/min the EDAC counts 38%, the Gampt 18% of total counts and at 6 l/min both the EDAC and Gampt only count 3% of total counts. CONCLUSIONS: Both the EDAC and Gampt can be used in a clinical setting for monitoring basal GME production. However, both devices have some major limitations when used for studying 'worst case' scenarios. One should take great caution when correlating measured data with neurocognitive outcome. Finally, results obtained by one device in a first study cannot be compared nor exchanged with results from the other device in a second study.
Authors:
Filip M J J De Somer; Maria R Vetrano; Jeroen P A J Van Beeck; Guido J Van Nooten
Publication Detail:
Type:  Comparative Study; Journal Article; Validation Studies     Date:  2010-03-02
Journal Detail:
Title:  Interactive cardiovascular and thoracic surgery     Volume:  10     ISSN:  1569-9285     ISO Abbreviation:  Interact Cardiovasc Thorac Surg     Publication Date:  2010 Jun 
Date Detail:
Created Date:  2010-05-18     Completed Date:  2010-08-24     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101158399     Medline TA:  Interact Cardiovasc Thorac Surg     Country:  England    
Other Details:
Languages:  eng     Pagination:  995-1001     Citation Subset:  IM    
Affiliation:
Heart Centre, Universitair Ziekenhuis Gent, Gent, Belgium. filip.desomer@ugent.be
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Blood Flow Velocity
Cardiac Surgical Procedures*
Cardiopulmonary Bypass / adverse effects*
Embolism, Air / etiology,  physiopathology,  ultrasonography*
Equipment Design
Humans
Lasers
Materials Testing
Microbubbles*
Monitoring, Intraoperative / instrumentation,  methods*
Predictive Value of Tests
Reproducibility of Results
Scattering, Radiation
Ultrasonography, Doppler, Pulsed* / instrumentation
Comments/Corrections
Comment In:
Interact Cardiovasc Thorac Surg. 2010 Jun;10(6):1001-2   [PMID:  20479078 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Short-term outcome following cardiac surgery - a comparison between consultant and trainees' perform...
Next Document:  Pulmonary valve repair late after right ventricular outflow tract reconstruction in children and ado...