Evaluation of six portable blood glucose meters for measuring blood glucose concentration in dogs.
MedLine Citation:
PMID:  19650698     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To evaluate accuracy of 6 portable blood glucose meters (PBGMs) by comparing results of these meters with results obtained with a reference chemistry analyzer. DESIGN: Evaluation study. ANIMALS: 49 dogs (158 blood samples). Procedures-Venous blood samples were tested with the 6 PBGMs, and results were compared with results of a commercially available analyzer that used a reference method based on the hexokinase reaction. RESULTS: Plasma glucose concentrations obtained with the reference analyzer ranged from 41 to 639 mg/dL. There were significant correlations between blood glucose concentrations obtained with the 6 PBGMs and plasma glucose concentrations obtained with the reference analyzer (r > or = 0.96). However, for all 6 PBGMs, results differed from results for the reference analyzer, with the difference increasing as plasma glucose concentration increased. Significant differences in bias were found among meters. For 142 samples classified as hypoglycemic, euglycemic, or hyperglycemic on the basis of results of the reference analyzer, the percentage of samples that were misclassified on the basis of results of the PBGMs ranged from 2.1% to 38.7%. CONCLUSIONS AND CLINICAL RELEVANCE: Results of the present study suggested that there were substantial differences in the accuracy of currently available PBGMs when used to determine blood glucose concentration in dogs.
Authors:
Todd A Cohen; Richard W Nelson; Philip H Kass; Mary M Christopher; Edward C Feldman
Related Documents :
2252758 - Development and evaluation of a wearable blood glucose monitor.
6691298 - Glucose-induced hemolysis of spheric red blood cells in hereditary spherocytosis: new a...
15048368 - Amperometric carbohydrate detection in flow systems by means of a bismuth modified plat...
7426808 - Local cerebral energy metabolism: its relationship to local functional activity and blo...
7364618 - Viscosity and density dependence during maximal flow in man.
2016458 - Relation between anterograde blood flow through a coronary artery and the size of the p...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Journal of the American Veterinary Medical Association     Volume:  235     ISSN:  0003-1488     ISO Abbreviation:  J. Am. Vet. Med. Assoc.     Publication Date:  2009 Aug 
Date Detail:
Created Date:  2009-08-04     Completed Date:  2009-09-25     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7503067     Medline TA:  J Am Vet Med Assoc     Country:  United States    
Other Details:
Languages:  eng     Pagination:  276-80     Citation Subset:  IM    
Affiliation:
Veterinary Medical Teaching Hospital, School of Veterinary Medicine, University of California, Davis, CA 95616, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Anticoagulants / pharmacology
Blood Glucose / analysis*
Blood Glucose Self-Monitoring / instrumentation,  standards,  veterinary*
Blood Specimen Collection / veterinary
Diabetes Mellitus / blood,  veterinary*
Dog Diseases / blood*
Dogs
Evaluation Studies as Topic
Female
Heparin / pharmacology
Male
Reference Values
Reproducibility of Results
Sensitivity and Specificity
Chemical
Reg. No./Substance:
0/Anticoagulants; 0/Blood Glucose; 9005-49-6/Heparin

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Heme Scavenging and the other facets of Hemopexin.
Next Document:  Incidence of and risk factors for postoperative pneumonia in dogs anesthetized for diagnosis or trea...